Bioelectrochemistry ; 133: 107448, 2020 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-31991387

RESUMO

We report on a novel impedance spectroscopy measurement and data analysis technique for cytotoxicity testing. The technique combines non-contact measurement with real-time impedance data analysis based on the toxin dose dependency of the outputs, making it suitable for high throughput screening. A multi-electrode array was designed and fabricated such that a standard well plate could be positioned above the electrodes, negating the requirement for bespoke culture wells with integrated electrodes. For cytotoxicity testing, endothelial cells, type ECV304, within the wells were exposed to various concentrations of 3 toxins, dimethyl sulphoxide, cadmium chloride and saponin, which exhibit different modes of action on cells. Impedance spectra were recorded every 30 min over a 24 h period. From the spectra 'toxin maps' were produced which presented the correlation between impedance output and dose of toxin versus frequency and time. The results demonstrated characteristic toxin maps for each toxin and significantly differences between the three toxins studied. Using complementary measurement methods, we showed that these differences in toxin maps related to morphological and physiological changes in the cells due to the differing mode of action of each toxin.


Assuntos
Técnicas Biossensoriais/instrumentação , Espectroscopia Dielétrica/instrumentação , Testes de Toxicidade/instrumentação , Linhagem Celular , Eletrodos , Células Endoteliais/efeitos dos fármacos , Desenho de Equipamento , Humanos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA