1.
Phys Med Biol ; 69(13)2024 Jun 26.
Artigo em Inglês | MEDLINE | ID: mdl-38862001

RESUMO

Objective.To present and characterize a novel method for x-ray computed tomography (xCT) calibration in proton treatment planning, based on proton CT (pCT) measurements on biological phantoms.Approach.A pCT apparatus was used to perform direct measurements of 3D stopping power relative to water (SPR) maps on stabilized, biological phantoms. Two single-energy xCT calibration curves-i.e. tissue substitutes and stoichiometric-were compared to pCT data. Moreover, a new calibration method based on these data was proposed, and verified against intra- and inter-species variability, dependence on stabilization, beam-hardening conditions, and analysis procedures.Main results.Biological phantoms were verified to be stable in time, with a dependence on temperature conditions, especially in the fat region: (-2.5 0.5) HU °C-1. The pCT measurements were compared with standard xCT calibrations, revealing an average SPR discrepancy within ±1.60% for both fat and muscle regions. In the bone region the xCT calibrations overestimated the pCT-measured SPR of the phantom, with a maximum discrepancy of about +3%. As a result, a new cross-calibration curve was directly extracted from the pCT data. Overall, the SPR uncertainty margin associated with this curve was below 3%; fluctuations in the uncertainty values were observed across the HU range. Cross-calibration curves obtained with phantoms made of different animal species and anatomical parts were reproducible with SPR discrepancies within 3%. Moreover, the stabilization procedure did not affect the resulting curve within a 2.2% SPR deviation. Finally, the cross-calibration curve was affected by the beam-hardening conditions on xCTs, especially in the bone region, while dependencies below 2% resulted from the image registration procedure.Significance.Our results showed that pCT measurements on biological phantoms may provide an accurate method for the verification of current xCT calibrations and may represent a tool for the implementation of a new calibration method for proton treatment planning.


Assuntos
Imagens de Fantasmas , Terapia com Prótons , Planejamento da Radioterapia Assistida por Computador , Tomografia Computadorizada por Raios X , Calibragem , Terapia com Prótons/instrumentação , Tomografia Computadorizada por Raios X/instrumentação , Planejamento da Radioterapia Assistida por Computador/métodos , Prótons , Animais , Humanos
2.
Phys Med Biol ; 69(17)2024 Aug 30.
Artigo em Inglês | MEDLINE | ID: mdl-39159669

RESUMO

Objective.Proton therapy administers a highly conformal dose to the tumour region, necessitating accurate prediction of the patient's 3D map of proton relative stopping power (RSP) compared to water. This remains challenging due to inaccuracies inherent in single-energy computed tomography (SECT) calibration. Recent advancements in spectral x-ray CT (xCT) and proton CT (pCT) have shown improved RSP estimation compared to traditional SECT methods. This study aims to provide the first comparison of the imaging and RSP estimation performance among dual-energy CT (DECT) and photon-counting CT (PCCT) scanners, and a pCT system prototype.Approach.Two phantoms were scanned with the three systems for their performance characterisation: a plastic phantom, filled with water and containing four plastic inserts and a wood insert, and a heterogeneous biological phantom, containing a formalin-stabilised bovine specimen. RSP maps were generated by converting CT numbers to RSP using a calibration based on low- and high-energy xCT images, while pCT utilised a distance-driven filtered back projection algorithm for RSP reconstruction. Spatial resolution, noise, and RSP accuracy were compared across the resulting images.Main results.All three systems exhibited similar spatial resolution of around 0.54 lp/mm for the plastic phantom. The PCCT images were less noisy than the DECT images at the same dose level. The lowest mean absolute percentage error (MAPE) of RSP,(0.28±0.07)%, was obtained with the pCT system, compared to MAPE values of(0.51±0.08)%and(0.80±0.08)%for the DECT- and PCCT-based methods, respectively. For the biological phantom, the xCT-based methods resulted in higher RSP values in most of the voxels compared to pCT.Significance.The pCT system yielded the most accurate estimation of RSP values for the plastic materials, and was thus used to benchmark the xCT calibration performance on the biological phantom. This study underlined the potential benefits and constraints of utilising such a novelex-vivophantom for inter-centre surveys in future.


Assuntos
Imagens de Fantasmas , Plásticos , Prótons , Tomografia Computadorizada por Raios X , Processamento de Imagem Assistida por Computador/métodos , Animais , Bovinos , Calibragem , Raios X
3.
Brain Behav ; 13(12): e3334, 2023 12.
Artigo em Inglês | MEDLINE | ID: mdl-38041516

RESUMO

INTRODUCTION: The purpose of the study is to investigate, by T2 relaxation, non-lesional white matter (WM) in relapsing-remitting (RR) multiple sclerosis (MS). METHODS: Twenty stable RR MS patients underwent 1.5T Magnetic Resonance Imaging (MRI) with 3D Fluid-Attenuated Inversion-Recovery (FLAIR), 3D-T1-weighted, and T2-relaxation multi-echo sequences. The Lesion Segmentation Tool processed FLAIR images to identify focal lesions (FLs), whereas T1 images were segmented to identify WM and FL sub-volumes with T1 hypo-intensity. Non-lesional WM was obtained as the segmented WM, excluding FL volumes. The multi-echo sequence allowed decomposition into myelin water, intra-extracellular water, and free water (Fw), which were evaluated on the segmented non-lesional WM. Correlation analysis was performed between the non-lesional WM relaxation parameters and Expanded Disability Status Scale (EDSS), disease duration, patient age, and T1 hypo-intense FL volumes. RESULTS: The T1 hypo-intense FL volumes correlated with EDSS. On the non-lesional WM, the median Fw correlated with EDSS, disease duration, age, and T1 hypo-intense FL volumes. Bivariate EDSS correlation of FL volumes and WM T2-relaxation parameters did not improve significance. CONCLUSION: T2 relaxation allowed identifying subtle WM alterations, which significantly correlated with EDSS, disease duration, and age but do not seem to be EDSS-predictors independent from FL sub-volumes in stable RR patients. Particularly, the increase in the Fw component is suggestive of an uninvestigated prodromal phenomenon in brain degeneration.


Assuntos
Esclerose Múltipla Recidivante-Remitente , Esclerose Múltipla , Substância Branca , Humanos , Lactente , Esclerose Múltipla Recidivante-Remitente/diagnóstico por imagem , Esclerose Múltipla Recidivante-Remitente/patologia , Substância Branca/diagnóstico por imagem , Substância Branca/patologia , Esclerose Múltipla/patologia , Imageamento por Ressonância Magnética/métodos , Água , Encéfalo/patologia
4.
Phys Med Biol ; 68(15)2023 07 24.
Artigo em Inglês | MEDLINE | ID: mdl-37379855

RESUMO

Objective. This paper describes the procedure to calibrate the three-dimensional (3D) proton stopping power relative to water (SPR) maps measured by the proton computed tomography (pCT) apparatus of the Istituto Nazionale di Fisica Nucleare (INFN, Italy). Measurements performed on water phantoms are used to validate the method. The calibration allowed for achieving measurement accuracy and reproducibility to levels below 1%.Approach. The INFN pCT system is made of a silicon tracker for proton trajectory determination followed by a YAG:Ce calorimeter for energy measurement. To perform the calibration, the apparatus has been exposed to protons of energies ranging from 83 to 210 MeV. Using the tracker, a position-dependent calibration has been implemented to keep the energy response uniform across the calorimeter. Moreover, correction algorithms have been developed to reconstruct the proton energy when this is shared in more than one crystal and to consider the energy loss in the non-uniform apparatus material. To verify the calibration and its reproducibility, water phantoms have been imaged with the pCT system during two data-taking sessions.Main results. The energy resolution of the pCT calorimeter resulted to beσEE≅0.9%at 196.5 MeV. The average values of the water SPR in fiducial volumes of the control phantoms have been calculated to be 0.995±0.002. The image non-uniformities were below 1%. No appreciable variation of the SPR and uniformity values between the two data-taking sessions could be identified.Significance. This work demonstrates the accuracy and reproducibility of the calibration of the INFN pCT system at a level below 1%. Moreover, the uniformity of the energy response keeps the image artifacts at a low level even in the presence of calorimeter segmentation and tracker material non-uniformities. The implemented calibration technique allows the INFN-pCT system to face applications where the precision of the SPR 3D maps is of paramount importance.


Assuntos
Terapia com Prótons , Prótons , Calibragem , Reprodutibilidade dos Testes , Tomografia Computadorizada por Raios X/métodos , Imagens de Fantasmas , Água , Terapia com Prótons/métodos
5.
Phys Med Biol ; 68(12)2023 06 07.
Artigo em Inglês | MEDLINE | ID: mdl-37201529

RESUMO

Objective. The goal of this study was to assess the imaging performances of the pCT system developed in the framework of INFN-funded (Italian National Institute of Nuclear Physics) research projects. The spatial resolution, noise power spectrum (NPS) and RSP accuracy has been investigated, as a preliminary step to implement a new cross-calibration method for x-ray CT (xCT).Approach. The INFN pCT apparatus, made of four planes of silicon micro-strip detectors and a YAG:Ce scintillating calorimeter, reconstructs 3D RSP maps by a filtered-back projection algorithm. The imaging performances (i.e. spatial resolution, NPS and RSP accuracy) of the pCT system were assessed on a custom-made phantom, made of plastic materials with different densities ((0.66, 2.18) g cm-3). For comparison, the same phantom was acquired with a clinical xCT system.Main results. The spatial resolution analysis revealed the nonlinearity of the imaging system, showing different imaging responses in air or water phantom background. Applying the Hann filter in the pCT reconstruction, it was possible to investigate the imaging potential of the system. Matching the spatial resolution value of the xCT (0.54 lp mm-1) and acquiring both with the same dose level (11.6 mGy), the pCT appeared to be less noisy than xCT, with an RSP standard deviation of 0.0063. Concerning the RSP accuracy, the measured mean absolute percentage errors were (0.23+-0.09)% in air and (0.21+-0.07)% in water.Significance. The obtained performances confirm that the INFN pCT system provides a very accurate RSP estimation, appearing to be a feasible clinical tool for verification and correction of xCT calibration in proton treatment planning.


Assuntos
Prótons , Tomografia Computadorizada por Raios X , Tomografia Computadorizada por Raios X/métodos , Raios X , Calibragem , Imagens de Fantasmas , Água