Anal Chem ; 86(16): 8344-51, 2014 Aug 19.
Artigo em Inglês | MEDLINE | ID: mdl-25135319

RESUMO

Current in vitro models of the leukocyte adhesion cascade cannot be used for real-time studies of the entire leukocyte adhesion cascade, including rolling, adhesion, and migration in a single assay. In this study, we have developed and validated a novel bioinspired microfluidic assay (bMFA) and used it to test the hypothesis that blocking of specific steps in the adhesion/migration cascade significantly affects other steps of the cascade. The bMFA consists of an endothelialized microvascular network in communication with a tissue compartment via a 3 µm porous barrier. Human neutrophils in bMFA preferentially adhered to activated human endothelial cells near bifurcations with rolling and adhesion patterns in close agreement with in vivo observations. Treating endothelial cells with monoclonal antibodies to E-selectin or ICAM-1 or treating neutrophils with wortmannin reduced rolling, adhesion, and migration of neutrophils to 60%, 20%, and 18% of their respective control values. Antibody blocking of specific steps in the adhesion/migration cascade (e.g., mAb to E-selectin) significantly downregulated other steps of the cascade (e.g., migration). This novel in vitro assay provides a realistic human cell based model for basic science studies, identification of new treatment targets, selection of pathways to target validation, and rapid screening of candidate agents.


Assuntos
Comunicação Celular , Endotélio Vascular/citologia , Leucócitos/citologia , Técnicas Analíticas Microfluídicas/métodos , Adulto , Adesão Celular , Células Endoteliais da Veia Umbilical Humana , Humanos , Migração e Rolagem de Leucócitos , Neutrófilos/citologia
2.
Biomed Opt Express ; 10(10): 5414-5430, 2019 Oct 01.
Artigo em Inglês | MEDLINE | ID: mdl-31646055

RESUMO

We have developed a human bronchial epithelial (HBE) cell and endothelial cell co-cultured microfluidic model to mimic the in vivo human airway. This airway-on-a-chip was designed with a central epithelial channel and two flanking endothelial channels, with a three-dimensional monolayers of cells growing along the four walls of the channel, forming central clear lumens. These cultures mimic airways and microvasculature in vivo. The central channel cells are grown at air-liquid interface and show features of airway differentiation including tight-junction formation, mucus production, and ciliated cells. Combined with novel micro-optical coherence tomography, this chip enables functional imaging of the interior of the lumen, which includes quantitation of cilia motion including beat frequency and mucociliary transport. This airway-on-a chip is a significant step forward in the development of microfluidics models for functional imaging.

3.
Sci Rep ; 8(1): 3171, 2018 02 16.
Artigo em Inglês | MEDLINE | ID: mdl-29453454

RESUMO

Assessment of anti-cancer drug efficacy in in vitro three-dimensional (3D) bioengineered cancer models provides important contextual and relevant information towards pre-clinical translation of potential drug candidates. However, currently established models fail to sufficiently recapitulate complex tumor heterogeneity. Here we present a chip-based tumor-mimetic platform incorporating a 3D in vitro breast cancer model with a tumor-mimetic microvascular network, replicating the pathophysiological architecture of native vascularized breast tumors. The microfluidic platform facilitated formation of mature, lumenized and flow-aligned endothelium under physiological flow recapitulating both high and low perfused tumor regions. Metastatic and non-metastatic breast cancer cells were maintained in long-term 3D co-culture with stromal fibroblasts in a poly(ethylene glycol)-fibrinogen hydrogel matrix within adjoining tissue chambers. The interstitial space between the chambers and endothelium contained pores to mimic the "leaky" vasculature found in vivo and facilitate cancer cell-endothelial cell communication. Microvascular pattern-dependent flow variations induced concentration gradients within the 3D tumor mass, leading to morphological tumor heterogeneity. Anti-cancer drugs displayed cell type- and flow pattern-dependent effects on cancer cell viability, viable tumor area and associated endothelial cytotoxicity. Overall, the developed microfluidic tumor-mimetic platform facilitates investigation of cancer-stromal-endothelial interactions and highlights the role of a fluidic, tumor-mimetic vascular network on anti-cancer drug delivery and efficacy for improved translation towards pre-clinical studies.


Assuntos
Antineoplásicos/farmacologia , Biomimética/instrumentação , Ensaios de Seleção de Medicamentos Antitumorais/instrumentação , Microvasos/efeitos dos fármacos , Desenho de Equipamento , Humanos , Dispositivos Lab-On-A-Chip , Células MCF-7 , Microvasos/fisiologia , Microambiente Tumoral/efeitos dos fármacos
4.
Anal Methods ; 7(4): 1273-1279, 2015 Feb 21.
Artigo em Inglês | MEDLINE | ID: mdl-26005497

RESUMO

This paper presents a bipolar electrode (BPE) device in a microfluidic dual-channel design for concurrent preconcentration and separation of composite DNA containing samples. The novelty of the present effort relies on the combination of BPE-induced ion concentration polarization (ICP) and end-labeled free-solution electrophoresis (ELFSE). The ion concentration polarization effect arising from the faradaic reaction on the BPE is utilized to exert opposing electrophoretic and electroosmotic forces on the DNA samples. Meanwhile, end-labeled free-solution electrophoresis alters the mass-charge ratio to enable simultaneous DNA separation in free solution. The microfluidic device was fabricated using standard and soft lithography techniques to form gold-on-glass electrode capped with a PDMS microfluidic channel. Experimental testing with various DNA samples was carried out over a range of applied electric field. Concentration ratios up to 285× within 5 minutes for a 102-mer DNA, and concurrent preconcentration and free-solution separation of binary mixture of free and bound 102-mer DNA within 6 minutes was demonstrated. The effect of applied electric field was also interrogated with respect to pertinent performance metrics of preconcentration and separation.

5.
Lab Chip ; 15(5): 1320-8, 2015 Mar 07.
Artigo em Inglês | MEDLINE | ID: mdl-25589423

RESUMO

This paper presents a continuous-flow microfluidic device for sorting stem cells and their differentiation progenies. The principle of the device is based on the accumulation of multiple dielectrophoresis (DEP) forces to deflect cells laterally in conjunction with the alternating on/off electric field to manipulate the cell trajectories. The microfluidic device containing a large array of oblique interdigitated electrodes was fabricated using a combination of standard and soft lithography techniques to generate a PDMS-gold electrode construct. Experimental testing with human mesenchymal stem cells (hMSC) and their differentiation progenies (osteoblasts) was carried out at different flow rates, and clear separation of the two populations was achieved. Most of the osteoblasts experiencing stronger DEP forces were deflected laterally and continuously, following zig-zag trajectories, and moved towards the desired collection outlet, whereas most of the hMSCs remained on the original trajectory due to weaker DEP forces. The experimental measurements were characterized and evaluated quantitatively, and consistent performance was demonstrated. Collection efficiency up to 92% and 67% for hMSCs and osteoblasts, respectively, along with purity up to 84% and 87% was obtained. The experimental results established the feasibility of our microfluidic DEP sorting device for continuous, label-free sorting of stem cells and their differentiation progenies.


Assuntos
Separação Celular/métodos , Eletroforese , Células-Tronco Mesenquimais/citologia , Técnicas Analíticas Microfluídicas/métodos , Diferenciação Celular , Separação Celular/instrumentação , Desenho de Equipamento , Humanos , Técnicas Analíticas Microfluídicas/instrumentação , Osteoblastos/citologia
6.
J Control Release ; 201: 49-55, 2015 Mar 10.
Artigo em Inglês | MEDLINE | ID: mdl-25599856

RESUMO

Tumor drug delivery is a complex phenomenon affected by several elements in addition to drug or delivery vehicle's physico-chemical properties. A key factor is tumor microvasculature with complex effects including convective transport, high interstitial pressure and enhanced vascular permeability due to the presence of "leaky vessels". Current in vitro models of the tumor microenvironment for evaluating drug delivery are oversimplified and, as a result, show poor correlation with in vivo performance. In this study, we report on the development of a novel microfluidic platform that models the tumor microenvironment more accurately, with physiologically and morphologically realistic microvasculature including endothelial cell lined leaky capillary vessels along with 3D solid tumors. Endothelial cells and 3D spheroids of cervical tumor cells were co-cultured in the networks. Drug vehicle screening was demonstrated using GFP gene delivery by different formulations of nanopolymers. The synthetic tumor network was successful in predicting in vivo delivery efficiencies of the drug vehicles. The developed assay will have critical applications both in basic research, where it can be used to develop next generation delivery vehicles, and in drug discovery where it can be used to study drug transport and delivery efficacy in realistic tumor microenvironment, thereby enabling drug compound and/or delivery vehicle screening.


Assuntos
Sistemas de Liberação de Medicamentos , Neoplasias/metabolismo , Microambiente Tumoral , Linhagem Celular , Técnicas de Cocultura , Células Endoteliais , Técnicas de Transferência de Genes , Proteínas de Fluorescência Verde/genética , Células HeLa , Humanos , Microfluídica , Nanopartículas/administração & dosagem , Nanopartículas/química , Plasmídeos , Polímeros/administração & dosagem , Polímeros/química
7.
Microfluid Nanofluidics ; 17(4): 693-699, 2014 Oct 01.
Artigo em Inglês | MEDLINE | ID: mdl-25346656

RESUMO

This paper presents a Nafion film based micro-nanofluidic device for concurrent DNA preconcentration and separation. The principle of the device is based on the combination of (a) ion concentration polarization phenomenon at the junction of the microchannel and the nanochannels in the Nafion film to form opposing electrophoretic and electroosmotic forces acting on the DNAs, and (b) end-labeled free solution electrophoresis to harness the charge-to-mass ratio for molecular differentiation. The experiments successfully demonstrated concurrent preconcentration and separation of DNA mixture in free solution within 240s, yielding concentration ratios up to 1,150X and separation resolution of 1.85. The effect of applied electric field on the concentration and separation performance was also investigated. The device can be used as a key sample preparation element in conjunction with micro- or nano-fluidic sensors for microTAS functionality.

8.
Lab Chip ; 13(12): 2300-10, 2013 Jun 21.
Artigo em Inglês | MEDLINE | ID: mdl-23636706

RESUMO

This paper presents a microfluidic electrical impedance flow cytometer (FC) for identifying the differentiation state of single stem cells. This device is comprised of a novel dual micropore design, which not only enhances the processing throughput, but also allows the associated electrodes to be used as a reference for one another. A signal processing algorithm, based on the support vector machine (SVM) theory, and a data classification method were developed to automate the identification of sample types and cell differentiation state based on measured impedance values. The device itself was fabricated using a combination of standard and soft lithography techniques to generate a PDMS-gold electrode construct. Experimental testing with non-biological particles and mouse embryonic carcinoma cells (P19, undifferentiated and differentiated) was carried out using a range of excitation frequencies. The effects of the frequency and the interrogation parameters on sample identification performance were investigated. It was found that the real and imaginary part of the detected impedance signal were adequate for distinguishing the undifferentiated P19 cells from non-biological polystyrene beads at all tested frequencies. A higher frequency and an opacity index were required to resolve the undifferentiated and differentiated P19 cells by capturing capacitive changes in electrophysiological properties arising from differentiation. The experimental results demonstrated salient accuracy of the device and algorithm, and established its feasibility for non-invasive, label-free identification of the differentiation state of the stem cells.


Assuntos
Técnicas Analíticas Microfluídicas/métodos , Células-Tronco Neoplásicas/citologia , Animais , Diferenciação Celular , Dimetilpolisiloxanos/química , Eletrodos , Citometria de Fluxo , Ouro/química , Camundongos , Técnicas Analíticas Microfluídicas/instrumentação , Poliestirenos/química , Máquina de Vetores de Suporte