1.
Acta Oncol ; 58(10): 1352-1357, 2019 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-31241387

RESUMO

Purpose: A 1.5 T MR Linac (MRL) has recently become available. MRL treatment workflows (WF) include online plan adaptation based on daily MR images (MRI). This study reports initial clinical experiences after five months of use in terms of patient compliance, cases, WF timings, and dosimetric accuracy. Method and materials: Two different WF were used dependent on the clinical situation of the day; Adapt To Position WF (ATP) where the reference plan position is adjusted rigidly to match the position of the targets and the OARs, and Adapt To Shape WF (ATS), where a new plan is created to match the anatomy of the day, using deformable image registration. Both WFs included three 3D MRI scans for plan adaptation, verification before beam on, and validation during IMRT delivery. Patient compliance and WF timings were recorded. Accuracy in dose delivery was assessed using a cylindrical diode phantom. Results: Nineteen patients have completed their treatment receiving a total of 176 fractions. Cases vary from prostate treatments (60Gy/20F) to SBRT treatments of lymph nodes (45 Gy/3F) and castration by ovarian irradiation (15 Gy/3F). The median session time (patient in to patient out) for 127 ATPs was 26 (21-78) min, four fractions lasted more than 45 min due to additional plan adaptation. For the 49 ATSs a median time of 12 (1-24) min was used for contouring resulting in a total median session time of 42 (29-91) min. Three SBRT fractions lasted more than an hour. The time on the MRL couch was well tolerated by the patients. The median gamma pass rate (2 mm,2% global max) for the adapted plans was 99.2 (93.4-100)%, showing good agreement between planned and delivered dose. Conclusion: MRL treatments, including daily MRIs, plan adaptation, and accurate dose delivery, are possible within a clinically acceptable timeframe and well tolerated by the patients.


Assuntos
Imageamento por Ressonância Magnética/métodos , Aceleradores de Partículas , Radiocirurgia/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Adulto , Idoso , Idoso de 80 Anos ou mais , Castração/instrumentação , Castração/métodos , Estudos de Coortes , Estudos de Viabilidade , Feminino , Humanos , Linfonodos/diagnóstico por imagem , Linfonodos/efeitos da radiação , Metástase Linfática/diagnóstico por imagem , Metástase Linfática/radioterapia , Imageamento por Ressonância Magnética/instrumentação , Masculino , Pessoa de Meia-Idade , Ovário/diagnóstico por imagem , Ovário/efeitos da radiação , Cooperação do Paciente/estatística & dados numéricos , Imagens de Fantasmas , Próstata/diagnóstico por imagem , Próstata/efeitos da radiação , Neoplasias da Próstata/diagnóstico por imagem , Neoplasias da Próstata/radioterapia , Radiometria , Radiocirurgia/instrumentação , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/instrumentação , Fatores de Tempo , Resultado do Tratamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA