RESUMO
Tunable nanopores fabricated in elastomeric membranes have been used to study the dependence of ionic current blockade rate on the concentration and electrophoretic mobility of particles in aqueous suspensions. A range of nanoparticle sizes, materials and surface functionalities has been tested. Using pressure-driven flow through a pore, the blockade rate for 100 nm carboxylated polystyrene particles was found to be linearly proportional to both transmembrane pressure (between 0 and 1.8 kPa) and particle concentration (between 7 × 10(8) and 4.5 × 10(10) ml( - 1)). This result can be accurately modelled using Nernst-Planck transport theory, enabling measurement of particle concentrations. Using only an applied potential across a pore, the blockade rates for carboxylic acid and amine coated 500 and 200 nm silica particles were found to correspond to changes in their mobility as a function of the solution pH. Scanning electron microscopy and confocal microscopy have been used to visualize changes in the tunable nanopore geometry in three dimensions as a function of applied mechanical strain. The pores were conical in shape, and changes in pore size were consistent with ionic current measurements. A zone of inelastic deformation adjacent to the pore has been identified as important in the tuning process.