Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Sensors (Basel) ; 23(17)2023 Aug 26.
Artigo em Inglês | MEDLINE | ID: mdl-37687894

RESUMO

A focused microwave thermotherapy system for non-invasively treating cancerous tumors has recently been actively developed. To accurately focus on the target location, the system needs information about the patient's body boundary. However, a water bolus is placed between the human body and the microwave applicators to allow the microwave to penetrate the body more easily and cool the body's skin. The structural configuration makes it difficult to measure the body boundary. In this paper, we present a body boundary measurement method using multiple line lasers and cameras for the application of a focused microwave thermotherapy system. Even with a lack of acquired boundary data, a completely closed boundary line can be reconstructed. In addition, real-time movement tracking is possible as it can be measured quickly, even in situations where the patient is moving, such as breathing and wriggling. The performance is verified with several indicators in a water-filled experimental testbed.


Assuntos
Hipertermia Induzida , Micro-Ondas , Humanos , Micro-Ondas/uso terapêutico , Estudo de Prova de Conceito , Lasers , Água
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA