Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Med Phys ; 27(2): 346-53, 2000 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-10718138

RESUMO

In this paper we present a complete description of the breathing synchronized radiotherapy (BSRT) system, which has been jointly developed between the University of California Davis Cancer Center and Varian Associates. BSRT is a description of an emerging radiation oncology procedure, where simulation, CT scan, treatment planning, and radiation treatment are synchronized with voluntary breath-hold, forced breath-hold, or breathing gating. The BSRT system consists of a breathing monitoring system (BMOS) and a linear accelerator gating hardware and software package. Two methods, a video camera-based method and the use of wraparound inductive plethysmography (RespiTrace), generate the BMOS signals. The BMOS signals and the synchronized fluoroscopic images are simultaneously recorded in the simulation room and are later analyzed to define the ideal treatment point (ITP) where organ motion is stationary. The BMOS signals at ITP can be used to gate a CT scanner or a linear accelerator to maintain the same organ configuration as in the simulation. The BSRT system allows breath-hold or gating. This dual role allows the system to be applicable for a variety of patients, i.e., the breath-hold method for those patients who can maintain and reproduce the ITP, and the forced breath-hold or gating method for those who are not ideal for voluntary breath-hold.


Assuntos
Radioterapia Assistida por Computador/métodos , Respiração , Fluoroscopia , Humanos , Processamento de Imagem Assistida por Computador , Monitorização Fisiológica , Aceleradores de Partículas , Pletismografia , Radioterapia Assistida por Computador/instrumentação , Tomografia Computadorizada por Raios X , Gravação em Vídeo
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA