Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 12 de 12
Filtrar
1.
Circulation ; 136(18): 1714-1725, 2017 Oct 31.
Artigo em Inglês | MEDLINE | ID: mdl-28674109

RESUMO

BACKGROUND: Reports of left ventricular assist device (LVAD) malfunction have focused on pump thrombosis. However, the device consists of the pump, driveline, and peripherals, all of which are potentially subject to failure. METHODS: Prospectively collected data were reviewed for all LVAD device malfunctions (DMs) occurring in rotary LVADs implanted at a single center between April 2004 and May 2016. Durable LVADs included 108 Heartmate II (HM II) and 105 HeartWare VAD (HVAD). DM data were categorized according to device type and into categories related to the component that failed: (1) controller, (2) peripheral components, and (3) implantable blood pump or its integral electric driveline. Pump-related events were analyzed as pump-specific (suspected or confirmed thrombosis) or nonpump-specific (driveline failure). DM rates were reported as events per 1000 patient-days, and Cox proportional hazard models were used for time-to-event analyses. Cumulative rates of malfunction were examined for the main components of each type of LVAD. RESULTS: Types of DM included controller failure (30%), battery failure (19%), or patient cable failure (14%), whereas only 13% were because of pump failure. DMs were more common in the HM II device (3.73 per 1000 patient-days versus 3.06 per 1000 patient-days for the HVAD, P<0.01). A higher rate of pump-specific malfunctions was discovered in those implanted with an HM II versus an HVAD (0.55 versus 0.39, respectively; P<0.01) and peripheral malfunctions (2.32 versus 1.78 for the HM II and HVAD, respectively; P<0.01); no difference occurred in the incidence of controller DM between the 2 LVADs. Patients with HVAD were 90% free of a pump-specific malfunction at 3 years compared with 56% for the HM II (log-rank P<0.003). Only 74% of the patients with HM II were free of pump thrombosis at 3 years compared with 90% of the patients with HVAD. Freedom from failure of the integrated driveline was 79% at 3 years for the HM II but 100% for the HVAD (log-rank P<0.02). CONCLUSIONS: Device malfunction is much broader than pump failure alone and occurs for different components at different rates based on the type of LVAD.


Assuntos
Análise de Falha de Equipamento , Coração Auxiliar/efeitos adversos , Trombose , Adulto , Idoso , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Estudos Prospectivos , Fatores de Risco , Trombose/epidemiologia , Trombose/etiologia
2.
Artif Organs ; 38(6): 447-55, 2014 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-24571597

RESUMO

Individual ventricular assist device (VAD) design may affect leukocytes and impact immunity. Few studies have presented leukocyte and infection profiles in VAD patients over the course of the implant period. CD11b (MAC-1) expression on granulocytes is an indicator of activation during inflammation, mediating extravasation and the release of reactive oxygen species in tissue. No reported studies have presented MAC-1 expression on circulating granulocytes in VAD patients. Fifty-six patients implanted at a single center with a HeartMate II (HMII; n = 32), HeartWare (HW; n = 12), or Thoratec pneumatic VAD (PVAD; n = 12) between 1999 and 2011 were followed for 120 days of support. The leukocyte profiles and infectious events of all patients were evaluated; additionally, a subset had MAC-1 expression on circulating granulocytes was measured (HMII n = 9; HW n = 7; PVAD n = 4). All groups exhibited a significant peak in leukocyte numbers at postoperative day (POD) 14 while simultaneously experiencing a significant decrease in hematocrit. HMII patients exhibited a 3.2-fold increase in granulocyte MAC-1 expression at POD 14, and the temporal trend over the implant period differed from that experienced by HW patients. Further, HW patients experienced significantly fewer infection events. Alterations in leukocyte profiles and granulocyte activation experienced by VAD patients appear to be device-specific. Elevations in leukocyte activation may be related to an increased risk for infection, although the specific relationship between these phenomena in this patient group is not known.


Assuntos
Granulócitos/imunologia , Insuficiência Cardíaca/terapia , Coração Auxiliar , Leucócitos/imunologia , Implantação de Prótese/instrumentação , Função Ventricular Esquerda , Adulto , Idoso , Biomarcadores/sangue , Feminino , Granulócitos/metabolismo , Insuficiência Cardíaca/diagnóstico , Insuficiência Cardíaca/fisiopatologia , Coração Auxiliar/efeitos adversos , Hematócrito , Humanos , Contagem de Leucócitos , Leucócitos/metabolismo , Antígeno de Macrófago 1/sangue , Masculino , Pessoa de Meia-Idade , Pennsylvania , Valor Preditivo dos Testes , Desenho de Prótese , Implantação de Prótese/efeitos adversos , Infecções Relacionadas à Prótese/sangue , Infecções Relacionadas à Prótese/diagnóstico , Infecções Relacionadas à Prótese/imunologia , Fatores de Risco , Infecção da Ferida Cirúrgica/sangue , Infecção da Ferida Cirúrgica/diagnóstico , Infecção da Ferida Cirúrgica/imunologia , Fatores de Tempo , Resultado do Tratamento
3.
Prog Transplant ; 23(1): 28-32, 2013 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-23448817

RESUMO

A patient with a HeartMate II left ventricular assist device who had a body mass index of 52 needed gastric bypass surgery in order to qualify for a heart transplant. Unlike previous experience in which the surgery was performed at the implant hospital, the gastric bypass surgery in this case was performed at a bariatric center of excellence that was a separate facility from the implant hospital. The artificial heart program of the University of Pittsburgh Medical Center worked with the bariatric center of excellence in scheduling the gastric bypass surgery using a multidisciplinary team approach at 2 hospitals to coordinate safe, high-quality patient care in a unique situation.


Assuntos
Derivação Gástrica , Coração Auxiliar , Planejamento de Assistência ao Paciente , Adulto , Insuficiência Cardíaca/cirurgia , Transplante de Coração , Humanos , Masculino , Obesidade Mórbida/cirurgia , Pennsylvania , Assistência Perioperatória
4.
Prog Transplant ; 22(4): 351-9, 2012 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-23187051

RESUMO

Ventricular assist devices (VADs) are a surgical treatment for heart failure. These devices may be implanted as a bridge to transplant or as destination therapy. After surgical recovery and education regarding device care, patients are discharged home. Meticulous care of the driveline must be taken to prevent infection and trauma of the site throughout the perioperative event and for the duration of support. Currently a standardized protocol for care of the driveline and exit site does not exist. VAD coordinators from across the country discussed the variability in care at different centers in the United States through a series of conference calls. A survey consisting of 16 questions was developed. The survey included questions on preoperative antibiotic recommendations, driveline placement and exit site suturing, frequency of dressing changes, and showering practices. VAD coordinators shared center-specific dressing protocols and any driveline success stories. This survey was sent to 73 centers; 38 centers (52%) responded. The purpose of the survey was to define current practice in order to move toward a standard of practice or protocol based on expert opinion for VAD driveline care and to assess the need for future studies.


Assuntos
Coração Auxiliar , Controle de Infecções/métodos , Infecções Relacionadas à Prótese/prevenção & controle , Antibioticoprofilaxia , Bandagens , Biofilmes , Humanos , Higiene , Hospedeiro Imunocomprometido , Infecções Relacionadas à Prótese/epidemiologia , Fatores de Risco , Inquéritos e Questionários , Técnicas de Sutura , Estados Unidos/epidemiologia
5.
Transplantation ; 105(3): 608-619, 2021 03 01.
Artigo em Inglês | MEDLINE | ID: mdl-32345866

RESUMO

BACKGROUND: Psychosocial evaluations are required for long-term mechanical circulatory support (MCS) candidates, no matter whether MCS will be destination therapy (DT) or a bridge to heart transplantation. Although guidelines specify psychosocial contraindications to MCS, there is no comprehensive examination of which psychosocial evaluation domains are most prognostic for clinical outcomes. We evaluated whether overall psychosocial risk, determined across all psychosocial domains, predicted outcomes, and which specific domains appeared responsible for any effects. METHODS: A single-site retrospective analysis was performed for adults receiving MCS between April 2004 and December 2017. Using an established rating system, we coded psychosocial evaluations to identify patients at low, moderate, or high overall risk. We similarly determined risk within each of 10 individual psychosocial domains. Multivariable analyses evaluated whether psychosocial risk predicted clinical decisions about MCS use (DT versus bridge), and postimplantation mortality, transplantation, rehospitalization, MCS pump exchange, and standardly defined adverse medical events (AEs). RESULTS: In 241 MCS recipients, greater overall psychosocial risk increased the likelihood of a DT decision (odds ratio, 1.76; P = 0.017); and postimplantation pump exchange and occurrence of AEs (hazard ratios [HRs] ≥ 1.25; P ≤ 0.042). The individual AEs most strongly predicted were cardiac arrhythmias and device malfunctions (HRs ≥ 1.39; P ≤ 0.032). The specific psychosocial domains predicting at least 1 study outcome were mental health problem severity, poorer medical adherence, and substance use (odds ratios and HRs ≥ 1.32; P ≤ 0.010). CONCLUSIONS: The psychosocial evaluation predicts not only clinical decisions about MCS use (DT versus bridge) but important postimplantation outcomes. Strategies to address psychosocial risk factors before or soon after implantation may help to reduce postimplantation clinical risks.


Assuntos
Insuficiência Cardíaca/terapia , Transplante de Coração/psicologia , Coração Auxiliar , Feminino , Seguimentos , Insuficiência Cardíaca/psicologia , Humanos , Masculino , Pessoa de Meia-Idade , Prognóstico , Estudos Retrospectivos , Fatores de Tempo
6.
Prog Transplant ; 20(2): 155-62, 2010 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-20642174

RESUMO

The Centers for Medicare and Medicaid Services announced that all hospitals implanting ventricular assist devices are required to have certification from the The Joint Commission for disease-specific care destination therapy with a ventricular assist device effective March 27, 2009, in order to receive Medicare reimbursement for services rendered to patients who have devices implanted for destination therapy. On February 23, 2007, The Joint Commission released the certification requirements for ventricular assist devices implanted for destination therapy in an 8-page document so that hospitals could prepare to meet the 2009 certification deadline. The Artificial Heart Program of the University of Pittsburgh Medical Center undertook a multidisciplinary project, under the guidance of the nurse coordinator, to prepare the hospital and program for a precertification survey by The Joint Commission for disease-specific destination therapy ventricular assist device certification. The Presbyterian Hospital Artificial Heart Program was awarded The Joint Commission's device-specific certification for destination therapy with ventricular assist devices in June 2008.


Assuntos
Institutos de Cardiologia/normas , Certificação , Coração Auxiliar , Implantação de Prótese/normas , Humanos , Capacitação em Serviço , Joint Commission on Accreditation of Healthcare Organizations , Medicare , Pennsylvania , Guias de Prática Clínica como Assunto , Desenvolvimento de Programas , Indicadores de Qualidade em Assistência à Saúde , Estados Unidos
7.
Prog Transplant ; 30(4): 376-381, 2020 12.
Artigo em Inglês | MEDLINE | ID: mdl-32985349

RESUMO

BACKGROUND: Ventricular assist device (VAD) patients are at high risk for morbidities and mortality. One potentially beneficial component of the Joint Commission VAD Certification process is the requirement that individual VAD programs select 4 performance measures to improve and optimize patients' clinical outcomes. PROBLEM STATEMENT: Review of patient data after our program's first certification visit in 2008 showed that, compared to national recommendations and published reports, our patients had suboptimal outcomes in 4 areas after device implantation: length of hospital stay, receipt of early (<48 hours) postsurgical physical therapy, driveline infection incidence, and adequacy of nutritional status (prealbumin ≥18 mg/dL). METHODS: Plan-Do-Study-Act processes were implemented to shorten length of stay, increase patient receipt of early physical therapy, decrease driveline infection incidence, and improve nutritional status. With 2008 as our baseline, we deployed interventions for each outcome area across 2009 to 2017. Performance improvement activities included staff, patient, and family didactic, one-on-one, and hands-on education; procedural changes; and outcomes monitoring with feedback to staff on progress. Descriptive and inferential statistics were examined to document change in the outcomes. OUTCOMES: Across the performance improvement period, length of stay decreased from 40 to 23 days; physical therapy consults increased from 87% to 100% of patients; 1-year driveline infection incidence went from 38% to 23.5%; and the percentage of patients with prealbumin within the normal range increased from 84% to 90%. IMPLICATIONS: Performance improvement interventions may enhance ventricular assist device patient outcomes. Interventions' sustainability should be evaluated to ensure that gains are not lost over time.


Assuntos
Procedimentos Cirúrgicos Cardíacos/normas , Coração Auxiliar/normas , Tempo de Internação/estatística & dados numéricos , Modalidades de Fisioterapia/normas , Guias de Prática Clínica como Assunto , Melhoria de Qualidade/normas , Disfunção Ventricular/cirurgia , Adulto , Idoso , Feminino , Humanos , Incidência , Masculino , Pessoa de Meia-Idade , Complicações Pós-Operatórias/prevenção & controle , Pré-Albumina/análise , Estudos Retrospectivos , Resultado do Tratamento , Estados Unidos
8.
Eur J Cardiothorac Surg ; 47(3): 497-504, 2015 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-24810756

RESUMO

OBJECTIVES: Preoperative liver dysfunction may influence haemostasis following ventricular assist device (VAD) implantation. The Model for End-stage Liver Disease (MELD) score was assessed as a predictor of bleeding and levels of haemostatic markers in patients with currently utilized VADs. METHODS: Sixty-three patients (31 HeartMate II, 15 HeartWare, 17 Thoratec paracorporeal ventricular assist device) implanted 2001-11 were analysed for preoperative liver dysfunction (MELD) and blood product administration. Of these patients, 21 had additional blood drawn to measure haemostatic marker levels. Cohorts were defined based on high (≥18.0, n = 7) and low (<18.0, n = 14) preoperative MELD scores. RESULTS: MELD score was positively correlated with postoperative administration of red blood cell (RBC), platelet, plasma and total blood product units (TBPU) , as well as chest tube drainage and cardiopulmonary bypass time. Age and MELD were preoperative predictors of TBPU by multivariate analysis. The high-MELD cohort had higher administration of TBPU, RBC and platelet units and chest tube drainage postimplant. Similarly, patients who experienced at least one bleeding adverse event were more likely to have had a high preoperative MELD. The high-MELD group exhibited different temporal trends in F1 + 2 levels and platelet counts to postoperative day (POD) 55. D-dimer levels in high-MELD patients became elevated versus those for low-MELD patients on POD 55. CONCLUSIONS: Preoperative MELD score predicts postoperative bleeding in contemporary VADs. Preoperative liver dysfunction may also alter postoperative subclinical haemostasis through different temporal trends of thrombin generation and platelet counts, as well as protracted fibrinolysis.


Assuntos
Transfusão de Sangue/métodos , Procedimentos Cirúrgicos Cardíacos/métodos , Insuficiência Cardíaca/fisiopatologia , Coração Auxiliar , Hepatopatias/fisiopatologia , Adulto , Idoso , Biomarcadores/sangue , Procedimentos Cirúrgicos Cardíacos/efeitos adversos , Estudos de Coortes , Feminino , Fibrinólise , Insuficiência Cardíaca/sangue , Insuficiência Cardíaca/terapia , Transplante de Coração , Hemodinâmica/fisiologia , Humanos , Hepatopatias/sangue , Masculino , Pessoa de Meia-Idade , Análise Multivariada , Contagem de Plaquetas
9.
Ann Thorac Surg ; 91(5): 1348-54, 2011 May.
Artigo em Inglês | MEDLINE | ID: mdl-21524442

RESUMO

BACKGROUND: Renal dysfunction is common before mechanical circulatory support (MCS). Mechanical circulatory support frequently improves renal function, but the impact of pre-MCS renal dysfunction on renal function after cardiac transplantation (CTX) is unknown. METHODS: Patients with MCS from January 1995 until April 2008 at a single center were included if their MCS duration was at least 60 days and they underwent successful CTX. Patients were followed for 1 year after CTX. RESULTS: A total of 116 patients were included in the study. Mechanical circulatory support was biventricular assist device in 28% and left ventricular assist device in 72% (continuous flow left ventricular assist device, 14%). Mean duration of MCS was 124 days. Patients were grouped according to tertiles of pre-MCS creatinine clearance (CrCl): group 1, CrCl less than 45 mL/min; group 2, CrCl between 45 and 65 mL/min inclusive; and group 3, CrCl more than 65 mL/min. Group 3 had the best renal outcomes both after MCS and 1 year after CTX. Regardless of group, patients who had a CrCl of at least 60 mL/min before CTX had similar 1-year posttransplant CrCl (55 versus 53 versus 56 mL/min for groups 1 through 3, respectively; not significantly different). However, the ability to achieve this level of renal function after MCS was less likely in those with the worst renal function before the initiation of MCS (53% versus 74% versus 90% for groups 1 through 3, respectively; p=0.001). CONCLUSIONS: The use of MCS leads to improvements in renal function in patients after MCS. However, the renal outcomes after CTX seem to be more dependent on the level of renal function achieved during MCS than on the level of renal function before MCS.


Assuntos
Insuficiência Cardíaca/diagnóstico , Insuficiência Cardíaca/cirurgia , Coração Auxiliar , Insuficiência Renal/diagnóstico , Análise de Variância , Estudos de Coortes , Creatinina/urina , Feminino , Seguimentos , Insuficiência Cardíaca/complicações , Transplante de Coração/métodos , Humanos , Testes de Função Renal , Masculino , Pessoa de Meia-Idade , Monitorização Fisiológica/métodos , Análise Multivariada , Cuidados Pós-Operatórios/métodos , Cuidados Pré-Operatórios/métodos , Valores de Referência , Insuficiência Renal/complicações , Estudos Retrospectivos , Medição de Risco , Fatores de Tempo , Resultado do Tratamento , Listas de Espera
10.
Ann Thorac Surg ; 92(4): 1420-6, 2011 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21958791

RESUMO

BACKGROUND: Obesity and heart failure are increasingly common, but the outcomes, weight changes, and adverse events of patients with advanced heart failure and obesity on mechanical support is not well described. METHODS: We retrospectively reviewed all non-underweight patients with durable mechanical support at a single institution from January 2000 until December 2008 and compared outcomes, weight change, and Interagency Registry for Mechanically Assisted Circulatory Support-defined adverse events between obese and nonobese patients. RESULTS: A total of 169 patients were included; 113 (67%) nonobese and 56 (33%) obese. Baseline characteristics, pump types, and implant duration were similar for both populations with the exception of more diabetes (61% vs 26%, p < 0.0001) and hypertension (61% vs 42%, p = 0.019) in the obese. Outcomes on mechanical support at 6 months were not different between groups. There was no significant difference between the nonobese and obese groups in the change in body mass index (-0.3 vs -1.0 mg/m(2), p = 0.29) over the duration of support. Obese patients, as compared with the nonobese, had higher incidence rates of sepsis (64.5% vs 34.7%, respectively, p = 0.006) and reoperation for infectious complications (34.2% vs 13.3%, respectively, p = 0.014). Obese patients also had a higher cumulative incidence of sepsis and reoperation for infection. Two-year posttransplant outcomes were not different in the obese and nonobese. CONCLUSIONS: Obese patients have similar outcomes on mechanical support, but at the cost of a higher cumulative incidence of sepsis and reoperations for infection; however, obese patients lose little weight while on mechanical support.


Assuntos
Insuficiência Cardíaca/terapia , Coração Auxiliar/efeitos adversos , Obesidade/complicações , Redução de Peso , Adulto , Idoso , Índice de Massa Corporal , Feminino , Seguimentos , Insuficiência Cardíaca/complicações , Insuficiência Cardíaca/mortalidade , Humanos , Masculino , Pessoa de Meia-Idade , Obesidade/mortalidade , Obesidade/fisiopatologia , Pennsylvania/epidemiologia , Prognóstico , Estudos Retrospectivos , Fatores de Risco , Taxa de Sobrevida/tendências , Resultado do Tratamento , Adulto Jovem
11.
J Heart Lung Transplant ; 29(9): 981-8, 2010 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-20580265

RESUMO

BACKGROUND: Ventricular assist devices (VADs) provide effective treatment for end-stage heart failure; however, most patients experience > or =1 major adverse events (AEs) while on VAD support. Although early, non-fatal AEs may increase the risk of later death during VAD support, this relationship has not been established. Therefore, we sought to determine the impact on 1-year mortality of AEs occurring during the first 60 days of VAD support. METHODS: A retrospective analysis was performed using prospectively collected data from a single-site database for patients aged > or =18 years receiving left ventricular or biventricular support during 1996 to 2008 and who survived >60 days on VAD support. Fourteen major classes of AEs occurring during this 60-day period were examined. One-year survival rates of patients with and without each major AE were compared. RESULTS: The study included 163 patients (80% men; mean age, 49.5 years), of whom 87% were European American, 72% had left ventricular support, and 83% were bridge to transplant. The occurrence of renal failure, respiratory failure, bleeding events, and reoperations during the first 60 days after implantation significantly increased the risk of 1-year mortality. After controlling for gender, age, VAD type, and intention to treat, renal failure was the only major AE significantly associated with later mortality (hazard ratio, 2.96; p = .023). CONCLUSIONS: Specific AEs, including renal failure, respiratory and bleeding events, and reoperations, significantly decrease longer-term survival. Renal failure conferred a 3-fold increased risk of 1-year mortality. Peri-operative management should focus on strategies to mitigate risk for renal failure in order to maximize later outcomes.


Assuntos
Insuficiência Cardíaca/cirurgia , Coração Auxiliar , Adulto , Idoso , Estudos de Coortes , Desenho de Equipamento , Feminino , Insuficiência Cardíaca/mortalidade , Transplante de Coração/mortalidade , Transplante de Coração/estatística & dados numéricos , Coração Auxiliar/efeitos adversos , Hemorragia/etiologia , Humanos , Infecções/etiologia , Nefropatias/etiologia , Masculino , Pessoa de Meia-Idade , Doenças Respiratórias/etiologia , Estudos Retrospectivos , Fatores de Risco , Taxa de Sobrevida , Falha de Tratamento , Resultado do Tratamento , Adulto Jovem
12.
Ann Thorac Surg ; 88(4): 1162-70, 2009 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-19766801

RESUMO

BACKGROUND: Although ventricular assist devices (VADs) provide effective treatment for end-stage heart failure, VAD support remains associated with significant risk for adverse events (AEs). To date there has been no detailed assessment of the incidence of a full range of AEs using standardized event definitions. We sought to characterize the frequency and timing of AE onset during the first 60 days of VAD support, a period during which clinical observation suggests the risk of incident AEs is high. METHODS: A retrospective analysis was performed utilizing prospectively collected data from a single-site clinical database including 195 patients aged 18 or greater receiving VADs between 1996 and 2006. Adverse events were coded using standardized criteria. Cumulative incidence rates were determined, controlling for competing risks (death, transplantation, recovery-wean). RESULTS: During the first 60 days after implantation, the most common AEs were bleeding, infection, and arrhythmias (cumulative incidence rates, 36% to 48%), followed by tamponade, respiratory events, reoperations, and neurologic events (24% to 31%). Other events (eg, hemolysis, renal, hepatic events) were less common (rates <15%). Some events (eg, bleeding, arrhythmias) showed steep onset rates early after implantation. Others (eg, infections, neurologic events) had gradual onsets during the 60-day period. Incidence of most events did not vary by implant era (1996 to 2000 vs 2001 to 2006) or by left ventricular versus biventricular support. CONCLUSIONS: Understanding differential temporal patterns of AE onset will allow preventive strategies to be targeted to the time periods when specific AE risks are greatest. The AE incidence rates provide benchmarks against which future studies of VAD-related risks may be compared.


Assuntos
Insuficiência Cardíaca/cirurgia , Coração Auxiliar , Complicações Pós-Operatórias/epidemiologia , Implantação de Prótese/efeitos adversos , Doença Aguda , Feminino , Humanos , Incidência , Masculino , Pessoa de Meia-Idade , Pennsylvania/epidemiologia , Complicações Pós-Operatórias/etiologia , Falha de Prótese , Estudos Retrospectivos , Fatores de Tempo
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA