An automated sample preparation module, based upon sequential injection analysis (SIA), has been developed for use within an autonomous pathogen detection system. The SIA system interfaced aerosol sampling with multiplexed microsphere immunoassay-flow cytometric detection. Metering and sequestering of microspheres using SIA was found to be reproducible and reliable, over 24-h periods of autonomous operation. Four inbuilt immunoassay controls showed excellent immunoassay and system stability over five days of unattended continuous operation. Titration curves for two biological warfare agents, Bacillus anthracis and Yersinia pestis, obtained using the automated SIA procedure were shown to be similar to those generated using a manual microtiter plate procedure.


Assuntos
Bacillus anthracis/química , Guerra Biológica , Monitoramento Ambiental/métodos , Yersinia pestis/química , Anticorpos Antibacterianos/imunologia , Bacillus anthracis/imunologia , Bacillus anthracis/patogenicidade , Monitoramento Ambiental/instrumentação , Citometria de Fluxo/instrumentação , Citometria de Fluxo/métodos , Imunoensaio/instrumentação , Imunoensaio/métodos , Microesferas , Fatores de Tempo , Yersinia pestis/imunologia , Yersinia pestis/patogenicidade