1.
J Chromatogr A ; 1457: 151-8, 2016 Jul 29.
Artigo em Inglês | MEDLINE | ID: mdl-27363736

RESUMO

The employment of an in-house-made capillary electrophoresis (CE) instrument with capacitively coupled contactless conductivity detection (C(4)D) as a simple and inexpensive solution for simultaneous determination of many rare earth elements (REEs) in ore samples from Vietnam, as well as in anti-corrosion coating samples is reported. 14 REEs (La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb and Lu) were determined using an electrolyte composed of 20mM arginine and 10mM α-hydroxyisobutyric acid adjusted to pH 4.2 with acetic acid. The best detection limit achieved was 0.24mg/L using the developed CE-C(4)D method. Good agreement between results from CE-C(4)D and the confirmation method (ICP-MS) was achieved, with a coefficient of determination (r(2)) for the two pairs of data of 0.998.


Assuntos
Metais Terras Raras/análise , Arginina , Corrosão , Eletrólitos , Eletroforese Capilar/métodos , Indústrias Extrativas e de Processamento , Concentração de Íons de Hidrogênio , Hidroxibutiratos , Limite de Detecção , Vietnã