Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Int J Oral Maxillofac Implants ; 10(2): 207-12, 1995.
Artigo em Inglês | MEDLINE | ID: mdl-7744440

RESUMO

Excessive heat generation at the implant-bone interface may cause irreversible bone damage and loss of osseointegration. The effect of heat generation in vitro at the implant surface caused by abutment reduction with medium- and extra-fine-grain diamond and tungsten burs in a high-speed dental turbine was examined. Titanium-alloy abutments connected to a titanium-alloy cylindrical implant embedded in an acrylic-resin mandible in a 37 degrees C water bath were reduced horizontally and vertically. Temperature changes were recorded via embedded thermocouples at the cervix and apex of the implant surface. Analysis of variance for repeated measures was used to compare seven treatment groups. Thirty seconds of continuous cutting with standard turbine coolant caused a mean temperature increase of 1 degrees C with a maximum of 2 degrees C. Similar tungsten cutting caused a mean increase of 2 degrees C with a maximum of 4.7 degrees C, significantly higher than diamond reduction. Additional air-water spray for continuous tungsten cutting had no significant effect, while intermittent cutting for 15-second increments reduced the temperature increase by 75%. Thus, abutment reduction with medium-grit diamonds using intermittent pressure and normal turbine coolant is unlikely to cause an interface-temperature increase sufficient to cause irreversible bone damage and compromise osseointegration.


Assuntos
Dente Suporte , Equipamentos Odontológicos de Alta Rotação , Implantação Dentária Endóssea , Implantes Dentários , Mandíbula/cirurgia , Resinas Acrílicas/química , Ar , Ligas , Análise de Variância , Ligas Dentárias/química , Diamante , Transferência de Energia , Temperatura Alta , Humanos , Mandíbula/fisiopatologia , Modelos Anatômicos , Osseointegração , Irrigação Terapêutica , Termômetros , Titânio/química , Tungstênio , Água
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA