Biomed Tech (Berl) ; 44(11): 300-7, 1999 Nov.
Artigo em Alemão | MEDLINE | ID: mdl-10608073

RESUMO

In this report, the results of complementary studies of pressure-volume (p-V) measurements on balloon catheters with balloons of low (LC) and high compliance (HC) used for percutaneous transluminal coronary angioplasty are discussed. The measurements were performed with balloons unconfined in air (free dilatation) and also confined in different shells. In the case of rigid shells, a surprisingly high self-expansion of the catheters was found. Although this self-expansion does not contribute to the radial dilatation, it cannot be neglected, but must be taken into account when the success of balloon dilatation is determined on the basis of measured p-V curves. The investigations performed using wrapped shells clearly show the different dilatation properties of LC and HC balloons. The results provide important information on the feasibility of controlled balloon dilatation on the basis of p-V measurements performed on-line during PTCA.


Assuntos
Angioplastia Coronária com Balão/instrumentação , Teste de Materiais , Complacência (Medida de Distensibilidade) , Desenho de Equipamento , Humanos , Modelos Cardiovasculares , Pressão