The global SARS-CoV-2 pandemic requires a rapid, reliable, and user-friendly diagnostic test to help control the spread of the virus. Reverse transcription and quantitative PCR (RT-qPCR) is currently the gold standard method for SARS-CoV-2 detection. Here, we develop a protocol based on reverse transcription loop-mediated isothermal amplification (RT-LAMP) and demonstrate increased sensitivity of this technique using fresh RNA extracts compared to RNA samples subjected to freezing/thawing cycles. We further compare RT-LAMP to RT-qPCR and demonstrate that the RT-LAMP approach has high sensitivity in fresh RNA extracts and can detect positive samples with Ct values between 8 and 35.


Assuntos
Teste de Ácido Nucleico para COVID-19/métodos , Técnicas de Diagnóstico Molecular/métodos , Técnicas de Amplificação de Ácido Nucleico/métodos , RNA Viral/isolamento & purificação , Colorimetria/métodos , Primers do DNA , Humanos , Nasofaringe/virologia , Transcrição Reversa , SARS-CoV-2/genética , Sensibilidade e Especificidade