RESUMO
PURPOSE OF REVIEW: Subclinical leaflet thrombosis (SLT) is often an incidental finding characterized by a thin layer of thrombus involving one, two or three leaflets, with typical appearance on multi-detector computed tomography (MDCT) of hypo-attenuating defect at the aortic side of the leaflet, also called hypo-attenuating leaflet thickening (HALT). SLT may occur following both transcatheter aortic replacement (TAVR) or biological surgical aortic valve replacement (SAVR). The aim of this review is to present an overview of the current state of knowledge on the incidence, diagnosis, clinical impact, and management of SLT following TAVR or SAVR. RECENT FINDINGS: SLT occurs in 10-20% of patients following TAVR and is somewhat more frequent than following SAVR (5-15%). SLT may regress spontaneously without treatment in about 50% of the cases but may also progress to clinically significant valve thrombosis in some cases. Oral anticoagulation with vitamin K antagonist is reasonable if SLT is detected by echocardiography and/or MDCT during follow-up and is generally efficient to reverse SLT. SLT is associated with mild increase in the risk of stroke but has no impact on survival. SLT has been linked with accelerated structural valve deterioration and may thus impact valve durability and long-term outcomes. SUMMARY: SLT is often an incidental finding on echocardiography or MDCT that occurs in 10-20% of patients following TAVR or 5-15% following biological SAVR and is associated with a mild increase in the risk of thrombo-embolic event with no significant impact on mortality but may be associated with reduced valve durability.
Assuntos
Valva Aórtica , Bioprótese , Próteses Valvulares Cardíacas , Trombose , Substituição da Valva Aórtica Transcateter , Humanos , Trombose/etiologia , Próteses Valvulares Cardíacas/efeitos adversos , Valva Aórtica/cirurgia , Bioprótese/efeitos adversos , Substituição da Valva Aórtica Transcateter/efeitos adversos , Complicações Pós-Operatórias/etiologia , Estenose da Valva Aórtica/cirurgia , Tomografia Computadorizada Multidetectores/métodos , Implante de Prótese de Valva Cardíaca/efeitos adversos , Implante de Prótese de Valva Cardíaca/métodosRESUMO
OBJECTIVES: The objective of this in vitro study was to compare Doppler versus catheter transvalvular pressure gradients (TPG) in third generations balloon-expandable (BE) versus self-expanding (SE) transcatheter heart valves (THV). BACKGROUND: TPG is a key parameter to assess and follow valve hemodynamic function following transcatheter aortic valve implantation (TAVI). It remains uncertain and debated whether, and to which extent, TPGs differ according to the type of THV, that is, BE versus SE and to the method used for TPG measurement, that is, Doppler echocardiography versus cardiac catheterization. METHODS: The CoreValve EVOLUT PRO 26 mm and the SAPIEN 3 23 mm THVs were tested in a left heart simulator using a 21 mm aortic annulus under following conditions: heart rate: 70 bpm, mean aortic pressure: 100 mmHg, stroke volume: 30, 70 and 100 ml. Mean TPGs were measured by continuous-wave Doppler and by micro-tip pressure catheters positioned in the left ventricle and at 50 mm downstream to the tip of the THV leaflets. RESULTS: Doppler TPGs (9.5 ± 3.9 mmHg) were on average 40.5 ± 13.9% higher (p < 0.001) than catheter TPGs (6.3 ± 3.4 mmHg). Both Doppler and catheter TPGs were lower (p = 0.003) in the SE versus BE THVs (Doppler: 8.7 ± 3.5 vs. 10.7 ± 4.6; catheter: 5.0 ± 1.7 mmHg vs. 7.1 ± 2.2). The Doppler versus catheter difference in TPG increased with the higher flow conditions. The Doppler versus catheter difference in TPG was similar in BE versus SE THVs (3.6 ± 1.1 vs. 3.7 ± 1.4 mmHg or 42 ± 9 vs. 47 ± 9%; p = 0.58) overall and in each flow conditions. CONCLUSION: The Doppler TPGs are, on average, 40% higher than the catheter TPGs for both BE and SE THVs. The SE THV had lower Doppler and catheter TPGs compared to the BE THV, at normal and high flow states. The absolute and percent differences between Doppler versus catheter TPGs were similar in BE versus SE THVs.
Assuntos
Estenose da Valva Aórtica , Próteses Valvulares Cardíacas , Substituição da Valva Aórtica Transcateter , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/cirurgia , Estenose da Valva Aórtica/diagnóstico por imagem , Estenose da Valva Aórtica/cirurgia , Catéteres , Humanos , Desenho de Prótese , Substituição da Valva Aórtica Transcateter/efeitos adversos , Resultado do TratamentoRESUMO
OBJECTIVES: Despite unusual high rates of patient-prosthesis mismatch (PPM), excellent midterm clinical outcomes have been reported after surgical aortic valve replacement (SAVR) with the Avalus™ bioprosthetic valve (Medtronic). To elucidate this "PPM conundrum," the Avalus valve haemodynamics were assessed during exercise testing. METHODS: Of the 148 patients who had undergone SAVR with the Avalus valve at our institution, 30 were randomly selected among those in whom stress test was deemed feasible and underwent a resting transthoracic echocardiography immediately followed by exercise echocardiography. Severe PPM was defined as indexed effective orifice area (iEOA) ≤ 0.65 cm2 /m2 and moderate PPM as iEOA > 0.65 and ≤ 0.85 cm2 /m2 . Measured PPM was determined with the use of the measured iEOA at rest or stress, while the estimated PPM was based on the estimated iEOA, derived from the mean EOA reported for each valve size in the manufacturer chart. RESULTS: Measured EOA significantly increased from rest to peak exercise in all PPM groups (p < .05) and the rates of moderate and severe measured PPM decreased from 40% and 20% to 27% and 0%, respectively. The patients with low-flow state (flow < 250 ml/s) had significantly lower measured rest EOA (p = .03). On the basis of the estimated iEOA, there was no severe PPM and 19 patients had moderate PPM (63.3%), with a significantly lower opening reserve than the patients without estimated PPM (p = .04). The estimated iEOA was more reliably correlated to the measured iEOA at maximal stress than the measured iEOA at rest, especially in patients with a low-flow state. CONCLUSIONS: This study supports the concept of an opening reserve of the Avalus valve to explain the PPM conundrum and promotes the use of exercise Doppler-echocardiography to complete the assessment of mismatch, especially in patients with a low-flow state. Published estimated EOA seems reliable to predict the haemodynamic performance of the Avalus valve, whether the flow conditions at rest.
Assuntos
Estenose da Valva Aórtica , Bioprótese , Implante de Prótese de Valva Cardíaca , Próteses Valvulares Cardíacas , Animais , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/cirurgia , Estenose da Valva Aórtica/diagnóstico por imagem , Estenose da Valva Aórtica/cirurgia , Bovinos , Hemodinâmica , Humanos , Desenho de Prótese , Resultado do TratamentoRESUMO
OBJECTIVE: The aim of this study was to assess the impact of sex on the management and outcome of patients according to aortic stenosis (AS) severity. INTRODUCTION: Sex differences in the management and outcome of AS are poorly understood. METHODS: Doppler echocardiography data of patients with at least mild-to-moderate AS [aortic valve area (AVA) ≤1.5 cm2 and peak jet velocity (VPeak) ≥2.5 m/s or mean gradient (MG) ≥25 mmHg] were prospectively collected between 2005 and 2015 and retrospectively analysed. Patients with reduced left ventricular ejection fraction (<50%), or mitral or aortic regurgitation >mild were excluded. RESULTS: Among 3632 patients, 42% were women. The mean indexed AVA (0.48 ± 0.17 cm2/m2), VPeak (3.74 ± 0.88 m/s), and MG (35.1 ± 18.2 mmHg) did not differ between sexes (all P ≥ 0.18). Women were older (72.9 ± 13.0 vs. 70.1 ± 11.8 years) and had more hypertension (75% vs. 70%; P = 0.0005) and less coronary artery disease (38% vs. 55%, P < 0.0001) compared to men. After inverse-propensity weighting (IPW), female sex was associated with higher mortality (IPW-HR: 1.91 [1.14-3.22]; P = 0.01) and less referral to valve intervention (competitive model IPW-HR: 0.88 [0.82-0.96]; P = 0.007) in the whole cohort. This excess mortality in women was blunted in concordant non-severe AS initially treated conservatively (IPW-HR = 1.03 [0.63-1.68]; P = 0.88) or in concordant severe AS initially treated by valve intervention (IPW-HR = 1.25 [0.71-2.21]; P = 0.43). Interestingly, the excess mortality in women was observed in discordant low-gradient AS patients (IPW-HR = 2.17 [1.19-3.95]; P = 0.01) where women were less referred to valve intervention (IPW-Sub-HR: 0.83 [0.73-0.95]; P = 0.009). CONCLUSION: In this large series of patients, despite similar baseline hemodynamic AS severity, women were less referred to AVR and had higher mortality. This seemed mostly to occur in the patient subset with discordant markers of AS severity (i.e. low-gradient AS) where women were less referred to AVR.
Assuntos
Estenose da Valva Aórtica , Implante de Prótese de Valva Cardíaca , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/cirurgia , Estenose da Valva Aórtica/diagnóstico por imagem , Estenose da Valva Aórtica/cirurgia , Feminino , Humanos , Masculino , Estudos Retrospectivos , Índice de Gravidade de Doença , Volume Sistólico , Resultado do Tratamento , Função Ventricular EsquerdaRESUMO
BACKGROUND: This study aimed to compare echocardiographic findings in low-risk patients with severe aortic stenosis after surgical aortic valve replacement (SAVR) or transcatheter aortic valve replacement (TAVR). METHODS: The PARTNER 3 trial (Placement of Aortic Transcatheter Valves) randomized 1000 patients with severe aortic stenosis and low surgical risk to undergo either transfemoral TAVR with the balloon-expandable SAPIEN 3 valve or SAVR. Transthoracic echocardiograms obtained at baseline and at 30 days and 1 year after the procedure were analyzed by a consortium of 2 echocardiography core laboratories. RESULTS: The percentage of moderate or severe aortic regurgitation (AR) was low and not statistically different between the TAVR and SAVR groups at 30 days (0.8% versus 0.2%; P=0.38). Mild AR was more frequent after TAVR than SAVR at 30 days (28.8% versus 4.2%; P<0.001). At 1 year, mean transvalvular gradient (13.7±5.6 versus 11.6±5.0 mm Hg; P=0.12) and aortic valve area (1.72±0.37 versus 1.76±0.42 cm2; P=0.12) were similar in TAVR and SAVR. The percentage of severe prosthesis-patient mismatch at 30 days was low and similar between TAVR and SAVR (4.6 versus 6.3%; P=0.30). Valvulo-arterial impedance (Zva), which reflects total left ventricular hemodynamic burden, was lower with TAVR than SAVR at 1 year (3.7±0.8 versus 3.9±0.9 mm Hg/mL/m2; P<0.001). Tricuspid annulus plane systolic excursion decreased and the percentage of moderate or severe tricuspid regurgitation increased from baseline to 1 year in SAVR but remained unchanged in TAVR. Irrespective of treatment arm, high Zva and low tricuspid annulus plane systolic excursion, but not moderate to severe AR or severe prosthesis-patient mismatch, were associated with increased risk of the composite end point of mortality, stroke, and rehospitalization at 1 year. CONCLUSIONS: In patients with severe aortic stenosis and low surgical risk, TAVR with the SAPIEN 3 valve was associated with similar percentage of moderate or severe AR compared with SAVR but higher percentage of mild AR. Transprosthetic gradients, valve areas, percentage of severe prosthesis-patient mismatch, and left ventricular mass regression were similar in TAVR and SAVR. SAVR was associated with significant deterioration of right ventricular systolic function and greater tricuspid regurgitation, which persisted at 1 year. High Zva and low tricuspid annulus plane systolic excursion were associated with worse outcome at 1 year whereas AR and severe prosthesis-patient mismatch were not. Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT02675114.
Assuntos
Estenose da Valva Aórtica/cirurgia , Valva Aórtica/cirurgia , Ecocardiografia , Implante de Prótese de Valva Cardíaca , Substituição da Valva Aórtica Transcateter , Idoso , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/fisiopatologia , Insuficiência da Valva Aórtica/diagnóstico por imagem , Insuficiência da Valva Aórtica/etiologia , Insuficiência da Valva Aórtica/fisiopatologia , Estenose da Valva Aórtica/diagnóstico por imagem , Estenose da Valva Aórtica/mortalidade , Estenose da Valva Aórtica/fisiopatologia , Feminino , Próteses Valvulares Cardíacas , Implante de Prótese de Valva Cardíaca/efeitos adversos , Implante de Prótese de Valva Cardíaca/instrumentação , Implante de Prótese de Valva Cardíaca/mortalidade , Hemodinâmica , Humanos , Masculino , América do Norte , Valor Preditivo dos Testes , Desenho de Prótese , Falha de Prótese , Medição de Risco , Fatores de Risco , Índice de Gravidade de Doença , Fatores de Tempo , Substituição da Valva Aórtica Transcateter/efeitos adversos , Substituição da Valva Aórtica Transcateter/instrumentação , Substituição da Valva Aórtica Transcateter/mortalidade , Resultado do TratamentoRESUMO
Infective endocarditis (IE) is a complex disease with cardiac involvement and multiorgan complications. Its prognosis depends on prompt diagnosis that leads to an aggressive therapeutic management combining antibiotic therapy and early cardiac surgery when indicated. However, IE diagnosis always poses a challenge, and echocardiography remains diagnostically imperfect in cases of prosthetic valve IE or cardiac implantable electronic device infection. In recent years, other imaging modalities (computed tomography, magnetic resonance imaging, nuclear imaging) have experienced significant technical improvements, and their application to the detection of cardiac and extracardiac IE-related lesions seems to be a strategic way forward in the management of patients with suspected IE. However, the scientific evidence in the literature remains limited; current guidelines address the use of the multimodality imaging in the field of IE with caution; the incremental value of each technique and their combinations is debated; and their use varies across countries. Despite these limitations, healthcare providers and surgeons should be aware of the possibilities offered by the multimodal imaging approach when appropriate. Here, we emphasize the value of a multidisciplinary heart valve team, the endocarditis team, underlining the importance of cardiac and extracardiac imaging experts in playing a key role in informing the diagnosis and management of patients with IE. Illustrative cases, critical appraisal of contemporary data, and conceptual and practical suggestions for clinicians that may help to improve the prognosis of patients with IE are provided in this review article.
Assuntos
Endocardite/diagnóstico por imagem , Imagem Multimodal , Equipe de Assistência ao Paciente , Comportamento Cooperativo , Endocardite/terapia , Humanos , Comunicação Interdisciplinar , Valor Preditivo dos Testes , Prognóstico , Reprodutibilidade dos TestesRESUMO
BACKGROUND: The incidence of structural valve deterioration after bioprosthesis (BP) aortic valve replacement (AVR) established on the basis of reoperation may substantially underestimate the true incidence. The objective is to determine the rate, timing, correlates, and association between hemodynamic valve deterioration (HVD) and outcomes assessed by Doppler echocardiography after surgical BP AVR. METHODS: A total of 1387 patients (62.2% male, 70.5±7.8 years of age) who underwent BP AVR were included in this retrospective study. Baseline echocardiography was performed at a median time of 4.1 (1.3-6.5) months after AVR. All patients had an echocardiographic follow-up ≥2 years after AVR (926 at least 5 years and 385 at least 10 years). HVD was defined by Doppler assessment as a ≥10 mm Hg increase in mean gradient or worsening of transprosthetic regurgitation ≥1/3 class. HVD was classified according to the timing after AVR: "very early," during the first 2-years; "early," between 2 and 5 years; "midterm," between 5 and 10 years; and "long-term," >10 years. RESULTS: A total of 428 patients (30.9%) developed HVD. Among these patients, 52 (12.0%) were classified as "very early," 129 (30.1%) as "early," 158 (36.9%) as "midterm," and 89 (20.8%) as "long-term" HVD. Factors independently associated with HVD occurring within the first 5 years after AVR were diabetes mellitus ( P=0.01), active smoking ( P=0.01), renal insufficiency ( P=0.01), baseline postoperative mean gradient ≥15 mm Hg ( P=0.04) or transprosthetic regurgitation ≥mild ( P=0.04), and type of BP (stented versus stentless, P=0.003). Factors associated with HVD occurring after the fifth year after AVR were female sex ( P=0.03), warfarin use ( P=0.007), and BP type ( P<0.001). HVD was independently associated with mortality (hazard ratio, 2.18; 95% CI, 1.86-2.57; P<0.001). CONCLUSIONS: HVD as identified by Doppler echocardiography occurred in one third of patients and was associated with a 2.2-fold higher adjusted mortality. Diabetes mellitus and renal insufficiency were associated with early HVD, whereas female sex, warfarin use, and stented BPs (versus stentless) were associated with late HVD.
Assuntos
Valva Aórtica/cirurgia , Bioprótese , Doenças das Valvas Cardíacas/cirurgia , Implante de Prótese de Valva Cardíaca/instrumentação , Próteses Valvulares Cardíacas , Hemodinâmica , Falha de Prótese , Idoso , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/fisiopatologia , Ecocardiografia Doppler , Feminino , Doenças das Valvas Cardíacas/diagnóstico por imagem , Doenças das Valvas Cardíacas/mortalidade , Doenças das Valvas Cardíacas/fisiopatologia , Implante de Prótese de Valva Cardíaca/efeitos adversos , Implante de Prótese de Valva Cardíaca/mortalidade , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese , Estudos Retrospectivos , Medição de Risco , Fatores de Risco , Fatores de Tempo , Resultado do TratamentoRESUMO
PURPOSE OF REVIEW: Bioprosthetic valves are now used for the majority of surgical aortic valve replacements and for all transcatheter aortic valve replacements. However, bioprostheses are subject to structural valve deterioration (SVD) and have, therefore limited durability. RECENT FINDINGS: Clinical, imaging, and circulating biomarkers may help to predict or indicate the presence of bioprosthetic valve SVD. The most important biomarkers of SVD includes: patient-related clinical biomarkers, such as diabetes and renal failure; valve-related biomarkers, such as absence of antimineralization process and severe prosthesis-patient mismatch; imaging biomarkers: the presence of valve leaflet mineralization on multidetector computed tomography or sodium fluoride uptake on positron emission tomography; and circulating biomarkers including: increased levels of HOMA index, ApoB/ApoA-I ratio, PCSK9, Lp-PLA2, phosphocalcic product. The assessment of these biomarkers may help to enhance risk stratification for SVD following AVR and may contribute to open novel pharmacotherapeutic avenues for the prevention of SVD. SUMMARY: SVD may affect all bioprostheses after aortic valve replacement, and is the main cause of bioprosthetic valve failure and reintervention during the follow-up. Comprehensive assessment of clinical, imaging, and circulating biomarkers associated with earlier SVD could help strengthen the follow-up in high-risk patients and provide novel pharmacologic therapeutic strategies.
Assuntos
Biomarcadores , Bioprótese , Implante de Prótese de Valva Cardíaca , Próteses Valvulares Cardíacas , Valva Aórtica , Biomarcadores/análise , Humanos , Pró-Proteína Convertase 9 , Desenho de Prótese , Falha de Prótese , Falha de TratamentoRESUMO
BACKGROUND: Left atrial appendage occlusion (LAAO) is increasingly used for stroke prevention in patients with atrial fibrillation who are considered unsuitable for a lifelong oral anticoagulant regimen. Recently, a single-centre study reported device-related thrombus formation in 16.7% of patients treated with the second-generation Amulet device (St. Jude Medical, St. Paul, MN, USA), presenting a potential major safety concern. As "real-world" data on device-related thrombus formation following LAAO with the Amulet occluder are scarce, we aimed to evaluate this outcome in a retrospective registry. METHODS: Clinical and tranosesophageal echocardiography data after LAAO with the Amulet in consecutive patients from three centres were collated. RESULTS: Among 38 patients (mean age 75.8 years), mean (standard deviation) CHA2DS2-VASc and HAS-BLED scores were 4.4 (1.2) and 3.4 (0.9), respectively. All patients underwent successful device placement without procedure-related adverse events. The antithrombotic regimen at discharge consisted of dual antiplatelet therapy (DAPT) in 27 patients (71.1%), single antiplatelet therapy in 10 patients (26.3%), and no antithrombotic therapy in one patient (2.6%). Device-related thrombus was observed in one patient (2.6%) despite DAPT regimen. The outcome of this patient was uncomplicated after adjustment of oral anticoagulant therapy. No patients presented with a thromboembolic event following LAAO during a mean (standard deviation) follow-up of 15 (5) months. CONCLUSIONS: In this retrospective study, device-related thrombus formation with the second-generation Amulet device was rare and occurred at a rate similar to that of the previous device. Importantly, no patient experienced a device-related thromboembolic event during follow-up. Larger real-life studies are required to confirm the safety profile of this increasingly used device.
Assuntos
Átrios do Coração , Cardiopatias/etiologia , Complicações Pós-Operatórias/etiologia , Dispositivo para Oclusão Septal/efeitos adversos , Acidente Vascular Cerebral/prevenção & controle , Trombose/etiologia , Administração Oral , Idoso , Idoso de 80 Anos ou mais , Anticoagulantes/administração & dosagem , Fibrilação Atrial/complicações , Fibrilação Atrial/terapia , Ecocardiografia Transesofagiana , Feminino , Seguimentos , Cardiopatias/diagnóstico , Humanos , Masculino , Pessoa de Meia-Idade , Complicações Pós-Operatórias/diagnóstico , Estudos Retrospectivos , Trombose/diagnósticoRESUMO
Aims: Leadless cardiac pacing has recently been proposed as alternative to conventional right ventricular (RV) pacing. With this approach, devices are directly screwed or fixed with tines in the RV wall, but the possible consequences on RV and tricuspid valve (TV) structure and function remain unknown. We thus conducted a study to evaluate this potential impact in chronically implanted patients. Methods and results: Repeated echocardiographic studies were performed prior to implantation, at discharge, and 2 months thereafter on all consecutive patients implanted with a leadless pacemaker at our centre between October 2014 and end-December 2015. Whenever possible, patients were evaluated in non-paced rhythm. Anatomical and functional parameters of RV, TV, and left cardiac structures were assessed. Overall, 23 patients (12 females, aged 85.2 ± 6.3 years) were included, with 14 implanted using Nanostim™ (Saint Jude Medical) and 9 with Micra™ (Medtronic). Indications for pacing were paroxysmal atrio-ventricular block in 12 patients, intermittent sinus bradycardia in 5, unexplained syncope in 3, and atrial fibrillation with slow ventricular rate in the remaining 3. The pacing percentage was 34 ± 42% at the last visit. Most devices were implanted in the septo-apical or mid-septal region. There were no significant changes in echocardiographic parameters observed. One patient developed significantly increased TV regurgitation, without abnormal leaflet motion or TV annulus size changes, suggesting it to be due to RV pressure changes. Conclusion: In patients chronically implanted with leadless pacemakers, there were no significant changes in heart structure and function observed, especially concerning the RV and TV.
Assuntos
Estimulação Cardíaca Artificial/efeitos adversos , Ventrículos do Coração , Implantação de Prótese , Insuficiência da Valva Tricúspide , Valva Tricúspide , Idoso , Idoso de 80 Anos ou mais , Ecocardiografia/métodos , Feminino , Seguimentos , França , Ventrículos do Coração/diagnóstico por imagem , Ventrículos do Coração/fisiopatologia , Humanos , Masculino , Implantação de Prótese/efeitos adversos , Implantação de Prótese/instrumentação , Implantação de Prótese/métodos , Valva Tricúspide/diagnóstico por imagem , Valva Tricúspide/fisiopatologia , Insuficiência da Valva Tricúspide/diagnóstico , Insuficiência da Valva Tricúspide/etiologiaRESUMO
PURPOSE: Ablation of premature ventricular complexes (PVCs) originating from left-sided papillary muscles is challenging. We tested a new approach by performing high-density mapping of PVC. METHODS AND RESULTS: We used a 20-pole deflectable spiral catheter during ablation procedures in four consecutive patients. Three presented with mitral valve prolapse, one with dilated cardiomyopathy. PVC burden was 24 ± 13%. The procedures lasted 182 ± 55.4 minutes, including 10 ± 3.2 minutes of radiofrequency. In all patients, mapping evidenced internal primary activation relative to the left ventricle shell (mean distance 21.3 ± 5.1 mm). Endocavitary prematurity was -38.3 ± 4.8 ms. Primary ablation success was achieved for all patients. CONCLUSIONS: High-density mapping of the papillary muscles in the left ventricle using a spiral catheter may be feasible. We identified the PVC foci away from the left ventricular shell. This consolidates the assumption for the origin of these ectopic beats at the junction between the chordae tendineae and the papillary muscles.
Assuntos
Músculos Papilares/cirurgia , Ablação por Radiofrequência/métodos , Complexos Ventriculares Prematuros/cirurgia , Cardiomiopatia Dilatada/fisiopatologia , Cardiomiopatia Dilatada/cirurgia , Ecocardiografia , Eletrocardiografia Ambulatorial , Mapeamento Epicárdico , Feminino , Fluoroscopia , Humanos , Masculino , Pessoa de Meia-Idade , Prolapso da Valva Mitral/fisiopatologia , Prolapso da Valva Mitral/cirurgia , Músculos Papilares/fisiopatologia , Resultado do Tratamento , Complexos Ventriculares Prematuros/fisiopatologiaRESUMO
OBJECTIVES: Much progress has been made in understanding the main causes of blood culture-negative endocarditis (BCNE). Few studies concerning BCNE treatment (due to previous antibiotics used or fastidious pathogens) are available. We performed this study to evaluate the effectiveness of our therapeutic protocol in BCNE, based on compliance with the protocol, outcome and 1 year mortality. PATIENTS AND METHODS: We collected prospectively and analysed retrospectively cases of BCNE between 2002 and 2014, using a simplified and standardized protocol developed by our multidisciplinary team. We apply two kinds of protocols to treat BCNE, which include only four intravenous antimicrobial agents: amoxicillin, vancomycin, gentamicin and amphotericin B. RESULTS: We had 177 patients with definite BCNE. There were 154 (87.0%) patients treated with both appropriate antimicrobial agents and appropriate duration of treatment. We analysed the causes of inappropriate treatment in 13 (7.3%) cases and inappropriate duration in 10 (5.6%) cases. The treatment changes were justified in all cases except one of discharge against medical advice. The fatality rate was 5.1% (nine cases) and all deaths occurred in the group of patients who were treated with appropriate treatment; however, four deaths were not attributable to empirical treatment failure. Concerning the other deaths, the lack of surgical management, in association with empirical treatment, could explain our protocol's failure, such as poorly tolerated surgery. CONCLUSIONS: Our protocol is efficient and our mortality rate was low, compared with the literature review. This may result from a strategy that uses a sampling procedure and a standardized protocol at the same time.
Assuntos
Anti-Infecciosos/administração & dosagem , Endocardite/tratamento farmacológico , Administração Intravenosa , Idoso , Endocardite/mortalidade , Feminino , Humanos , Masculino , Estudos Prospectivos , Estudos Retrospectivos , Análise de Sobrevida , Resultado do TratamentoRESUMO
Antiphospholipid syndrome (APS) is an autoimmune disorder with combination of at least 1 clinical and 1 laboratory criterion as defined by the SAPPORO statement. Clinical criteria result from vascular thrombosis that can affect artery, venous, or small vessel in any tissue or organ. Arterial stenosis is a rare lesion involved in APS, affecting mainly renal or intracranial arteries. We reported a case of a 33-year-old woman with abdominal angina and high blood pressure (BP). Imaging showed tight, not calcified, and hypodense stenosis of mesenteric superior artery and left renal artery, and a thrombosis of the celiac trunk. Treatment was digestive rest followed by angioplasty and stenting of mesenteric and renal artery, anticoagulation, antiplatelet, and statin therapy. Normal BP and digestive function were obtained postoperatively. Biological tests showed a positive lupus anticoagulant at diagnosis and at 12 weeks, which allowed us to make the diagnosis of APS. Physiopathology of stenosis in APS remains unclear but suggests arterial wall partial thrombosis, accelerated atherosclerosis, and/or proliferation of smooth muscle cells. We recommend screening of arterial stenosis in patients with APS and arterial symptoms, and inversely, searching for APS in young patients with atypical arterial stenosis to allow optimal therapy.
Assuntos
Arteriopatias Oclusivas/cirurgia , Isquemia/etiologia , Perna (Membro)/irrigação sanguínea , Artéria Poplítea , Feminino , Humanos , MasculinoRESUMO
Transcatheter aortic valve implantation (TAVI) is now utilised as a less invasive alternative to surgical aortic valve replacement (SAVR) across the whole spectrum of surgical risk. Long-term durability of the bioprosthetic valves has become a key goal of TAVI as this procedure is now considered for younger and lower-risk populations. The purpose of this article is to present a state-of-the-art overview on the definition, aetiology, risk factors, mechanisms, diagnosis, clinical impact, and management of bioprosthetic valve dysfunction (BVD) and failure (BVF) following TAVI with a comparative perspective versus SAVR. Structural valve deterioration (SVD) is the main factor limiting the durability of the bioprosthetic valves used for TAVI or SAVR, but non-structural BVD, such as prosthesis-patient mismatch and paravalvular regurgitation, as well as valve thrombosis or endocarditis may also lead to BVF. The incidence of BVF related to SVD or other causes is low (<5%) at midterm (5- to 8-year) follow-up and compares favourably with that of SAVR. The long-term follow-up data of randomised trials conducted with the first generations of transcatheter heart valves also suggest similar valve durability in TAVI versus SAVR at 10 years, but these trials suffer from major survivorship bias, and the long-term durability of TAVI will need to be confirmed by the analysis of the low-risk TAVI versus SAVR trials at 10 years.
Assuntos
Valva Aórtica , Próteses Valvulares Cardíacas , Substituição da Valva Aórtica Transcateter , Humanos , Substituição da Valva Aórtica Transcateter/efeitos adversos , Substituição da Valva Aórtica Transcateter/instrumentação , Substituição da Valva Aórtica Transcateter/métodos , Valva Aórtica/cirurgia , Valva Aórtica/diagnóstico por imagem , Fatores de Risco , Bioprótese , Estenose da Valva Aórtica/cirurgia , Falha de Prótese , Resultado do Tratamento , Desenho de Prótese , Implante de Prótese de Valva Cardíaca/métodos , Implante de Prótese de Valva Cardíaca/instrumentação , Implante de Prótese de Valva Cardíaca/efeitos adversosRESUMO
Background: It is unknown whether bioprostheses used for transcatheter aortic valve implantation will have similar long-term durability as those used for surgical aortic valve replacement. Repetitive mechanical stress applied to the valve leaflets, particularly during diastole, is the main determinant of structural valve deterioration. Leaflet mechanical stress cannot be measured in vivo. The objective of this in vitro/in silico study was thus to compare the magnitude and regional distribution of leaflet mechanical stress in old vs new generations of self-expanding (SE) vs balloon expandable (BE) transcatheter heart valves (THVs). Methods: A double activation simulator was used for in vitro testing of two generations of SE THV (Medtronic CoreValve 26 mm and EVOLUT PRO 26 mm) and two generations of BE THV (Edwards SAPIEN 23 mm vs SAPIEN-3 23 mm). These THVs were implanted within a 21-mm aortic annulus. A noncontact system based on stereophotogammetry and digital image correlation with high spatial and temporal resolution (2000 img/sec) was used to visualize the valve leaflet motion and perform the three-dimensional analysis. A finite element model of the valve was developed, and the leaflet deformation obtained from the digital image correlation analysis was applied to the finite element model to calculate local leaflet mechanical stress during diastole. Results: The maximum von Mises leaflet stress was higher in early vs new THV generation (p < 0.05) and in BE vs SE THV (p < 0.05): early generation BE: 2.48 vs SE: 1.40 MPa; new generation BE: 1.68 vs SE: 1.07 MPa. For both types of THV, the highest values of leaflet stress were primarily observed in the upper leaflet edge near the commissures and to a lesser extent in the mid-portion of the leaflet body, which is the area where structural leaflet deterioration most often occurs in vivo. Conclusions: The results of this in vitro/in silico study suggest that: i) Newer generations of THVs have â¼30% lower leaflet mechanical stress than the early generations; ii) For a given generation, SE THVs have lower leaflet mechanical stress than BE THVs. Further studies are needed to determine if these differences between new vs early THV generations and between SE vs BE THVs will translate into significant differences in long-term valve durability in vivo.
RESUMO
Background: Residual mitral regurgitation (MR) is associated with worse outcomes after transcatheter edge-to-edge mitral valve repair (TEER). Shear stress induced by MR leads to altered von Willebrand factor activity (vWF:Act) and increased closure time with adenosine diphosphate (CT-ADP). Objectives: The purpose of this study was to investigate the use of CT-ADP to monitor MR during TEER and the association between the vWF, residual MR, and clinical events post-TEER. Methods: Sixty-five patients undergoing TEER were enrolled. CT-ADP was measured at baseline, after each clip deployment, 1 hour and 24 hours post-TEER. CT-ADP values were related to vWF:Act/vWF antigen (vWF:Ag) ratio at the same time points, and MR severity was assessed by echocardiography at 1 month. Combined events of all-cause mortality and heart failure hospitalizations were evaluated at 1 year. Results: At 1 month, 32 (49%) patients had residual MR > mild (of those, 14% had MR > moderate). There was no significant change in CT-ADP values during the procedure. However, CT-ADP significantly decreased 1-hour post-TEER (P < 0.001). Patients with corrected MR demonstrated an increase in vWF:Act/vWF:Ag ratio 1-hour post-TEER. Elevated baseline vWF:Act/vWF:Ag ratio and the periprocedural percentage changes of the vWF:Act/vWF:Ag ratio (1 hour post-TEER - baseline values) were associated with the combined clinical outcome. Conclusions: CT-ADP evolution in time was not quick enough to provide real-time monitoring of MR severity during TEER. However, vWF:Act/vWF:Ag ratio at baseline and its variations following the procedure were associated with clinical outcomes. Those findings will need external validation.
RESUMO
BACKGROUND: Bicuspid aortic valve (BAV) is often associated with a concomitant aortopathy. However, few studies have evaluated the effect of the aortic valve (AV) phenotype on the rate of dilation of the aorta. This study aimed to compare the progression rate of aorta dimensions according to AV phenotype (BAV vs tricuspid AV (TAV)), fusion type and sex in patients with aortic stenosis (AS). METHODS: 310 patients with AS (224 TAV and 86 BAV) recruited in the Metabolic Determinants of the Progression of Aortic Stenosis study (PROGRESSA, NCT01679431) were included in this analysis. Doppler echocardiography was performed annually to assess AS severity and measure ascending aorta (AA) dimensions. Baseline and last follow-up visit measurements were used to assess the annualised change. RESULTS: Median AA annualised change was larger in BAV versus TAV (0.33±0.65 mm/year vs 0.21±0.56 mm/year, p=0.04). In the whole cohort, BAV phenotype and higher low-density lipoprotein (LDL) levels were significantly associated with fast progression of AA dilation in univariate analysis (OR 1.80, 95% CI 1.08 to 2.98, p=0.02; 1.37, 95% CI 1.04 to 1.80, p=0.03, respectively). AA dilation rate did not vary according to the BAV subtype (p=0.142). Predictors of AA progression rate were different between valve phenotypes, with higher apolipoprotein B/apolipoprotein A-I ratio, higher baseline peak aortic jet velocity (Vpeak) and smaller baseline AA diameter in the TAV cohort (all p<0.05) versus absence of hypertension, higher LDL levels and smaller baseline AA diameter in the BAV cohort (all p<0.02). In men, higher baseline Vpeak and smaller baseline AA (p<0.001) were independently associated with increased annualised AA dilation, while in women, higher LDL levels (p=0.026) were independently associated with faster AA dilation. CONCLUSION: This study suggests that BAV is associated with faster dilation of the AA. Predictors of AA dilation are different between valve phenotype and sex, with higher LDL levels being associated with faster AA dilation in BAV.
Assuntos
Estenose da Valva Aórtica , Valva Aórtica , Doença da Válvula Aórtica Bicúspide , Progressão da Doença , Ecocardiografia Doppler , Fenótipo , Humanos , Masculino , Estenose da Valva Aórtica/fisiopatologia , Estenose da Valva Aórtica/diagnóstico , Estenose da Valva Aórtica/complicações , Feminino , Idoso , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/anormalidades , Valva Aórtica/patologia , Valva Aórtica/fisiopatologia , Fatores Sexuais , Ecocardiografia Doppler/métodos , Doença da Válvula Aórtica Bicúspide/fisiopatologia , Dilatação Patológica , Seguimentos , Pessoa de Meia-Idade , Estudos Prospectivos , Fatores de Risco , Aorta/diagnóstico por imagem , Aorta/fisiopatologia , Idoso de 80 Anos ou mais , Índice de Gravidade de DoençaRESUMO
Cardiac troponin is extensively used as a biomarker in modern medicine due to its diagnostic capability for myocardial injury, as well as its predictive and prognostic value for cardiac diseases. However, heterophile antibodies, antitroponin antibodies, and macrotroponin complexes can be observed both in seemingly healthy individuals and patients with cardiac diseases, potentially leading to false positive or disproportionate elevation of cTn (cardiac troponin) assay results and introducing discrepancies in clinical interpretations with impact on medical management. In this review article, we describe the possible mechanisms of cTn release and the sources of variations in the assessment of circulating cTn levels. We also explore the pathophysiological mechanisms underlying antitroponin antibody development and discuss the influence exerted by macrotroponin complexes on the results of immunoassays. Additionally, we explore approaches to detect these complexes by presenting various clinical scenarios encountered in routine clinical practice. Finally, unsolved questions about the development, prevalence, and clinical significance of cardiac autoantibodies are discussed.
Assuntos
Autoanticorpos , Biomarcadores , Humanos , Biomarcadores/sangue , Autoanticorpos/sangue , Cardiopatias/diagnóstico , Cardiopatias/sangue , Cardiopatias/imunologia , Valor Preditivo dos Testes , Troponina I/sangue , Troponina I/imunologia , PrognósticoRESUMO
AIMS: Evaluation of left and right ventricular (RV) longitudinal systolic function may enhance risk stratification following aortic valve replacement (AVR). The study objective was to evaluate the changes in left and RV longitudinal systolic function and RV-pulmonary artery (RV-PA) coupling from baseline to 30 days and 1 year after AVR. METHODS AND RESULTS: Left ventricular (LV) longitudinal strain (LS), tricuspid annulus plane systolic excursion (TAPSE), and RV-PA coupling were evaluated in patients from the PARTNER 2A surgical AVR (SAVR) arm (n = 985) and from the PARTNER 2 SAPIEN 3 registry (n = 719). TAPSE and RV-PA coupling decreased significantly following SAVR, but remained stable following TAVR. Lower LV LS, TAPSE, or RV-PA coupling at baseline was associated with increased risk of the composite of death, hospitalization, and stroke at 5 years [adjusted hazard ratios (HRs) for LV LS < 15%: 1.24, 95% confidence interval (CI) 1.05-1.45, P = 0.001; TAPSE < 14â mm: 1.44, 95% CI 1.21-1.73, P < 0.001; RV-PA coupling < 0.55â mm/mmHg: 1.32, 95% CI 1.07-1.63, P = 0.011]. Reduced TAPSE at baseline was the most powerful predictor of the composite endpoint at 5 years. Patients with LV ejection fraction <50% at baseline had increased risk of the primary endpoint with SAVR (HR: 1.34, 95% CI 1.08-1.68, P = 0.009) but not with TAVR (HR: 1.12, 95% CI 0.88-1.42). Lower RV-PA coupling at 30 days showed the strongest association with cardiac mortality. CONCLUSION: SAVR but not TAVR was associated with a marked deterioration in RV longitudinal systolic function and RV-PA coupling. Lower TAPSE and RV-PA coupling at 30 days were associated with inferior clinical outcomes at 5 years. In patients with LVEF < 50%, TAVR was associated with superior 5-year outcomes.