RESUMO
A high-performance liquid chromatography-tandem mass spectrometry (LC/MS/MS) technique was developed for the simultaneous determination of five non-steroidal anti-inflammatory oxicam drugs (ampiroxicam, tenoxicam, piroxicam, meloxicam and lornoxicam) in human plasma. These five oxicam drugs and isoxicam (internal standard) were extracted from human plasma with an Oasis(®) MAX cartridge column and analysed on a Unison UK-C18 column (2.0 mm × 100 mm, 3 µm) with an acetonitrile:10mM formic ammonium buffer (pH 3.0) (50:50) mobile phase at 0.20 ml/min at 37°C. The analytes were detected using a tandem mass spectrometer, equipped with an electrospray ion source (ESI). The instrument was used in multiple-reaction-monitoring (MRM) mode. The extraction yields from a 200 µl human plasma sample (containing 10 ng of each drugs) with the Oasis(®) MAX cartridge column were 93.3-102.5%. The detection limits were 0.01-6.5 ng/ml (S/N=3). Our developed method is very useful for the simultaneous determination of five oxicam (non-steroidal anti-inflammatory) drugs in human plasma by LC/MS/MS.