RESUMO
Integral microscopes (IMic) have been recently developed in order to capture the spatial and the angular information of 3D microscopic samples with a single exposure. Computational post-processing of this information permits to carry out a 3D reconstruction of the sample. By applying conventional algorithms, both depth and also view reconstructions are possible. However, the main drawback of IMic is that the resolution of the reconstructed images is low and axially heterogeneous. In this paper, we propose a new configuration of the IMic by placing the lens array not at the image plane, but at the pupil (or Fourier) plane of the microscope objective. With this novel system, the spatial resolution is increased by factor 1.4, and the depth of field is substantially enlarged. Our experiments show the feasibility of the proposed method.
RESUMO
In this work, Fourier integral microscope (FIMic), an ultimate design of 3D-integral microscopy, is presented. By placing a multiplexing microlens array at the aperture stop of the microscope objective of the host microscope, FIMic shows extended depth of field and enhanced lateral resolution in comparison with regular integral microscopy. As FIMic directly produces a set of orthographic views of the 3D-micrometer-sized sample, it is suitable for real-time imaging. Following regular integral-imaging reconstruction algorithms, a 2.75-fold enhanced depth of field and [Formula: see text]-time better spatial resolution in comparison with conventional integral microscopy is reported. Our claims are supported by theoretical analysis and experimental images of a resolution test target, cotton fibers, and in-vivo 3D-imaging of biological specimens.