RESUMO
A Fluorescence probe was designed based on 8-hydroxyquinoline chitosan silica precursor (HQCS) for selective detection of Al3+, Cu2+. The HQCS has no observable fluorescence signal, but after the addition of Al3+, a huge fluorescence signal appeared, and the selective quenching was absorbed after the addition of Cu2+. The effect of other different cations, including Cu2+, Mg2+, Ca2+, Pb2+, Zn2+, Hg2+, Ag+, Fe3+, and K+ was studied. The addition of Cu2+ to the probe (HQCSAL) decreased the fluorescence very repeatable, and the variation of the fluorescence vs. Cu2+ was monotonic and linear. Therefore, the prepared probe was used to determine Cu2+ ions in real samples. The mechanism of fluorescence variation by adding cations to the probe solution was studied using the Stern-Volmer equation. Under the optimum conditions, the linear range and detection limit were 3.5-31 µM and 1 µM, respectively. The probe accuracy on the copper determination in the blood and tap waters was comparable to the ICP-OES results. The circuit logic gate mimic was designed for the fluorescence behavior of the probe constituents.