Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Innovations (Phila) ; 6(4): 253-6, 2011 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-22437984

RESUMO

OBJECTIVE: The placement of epicardial pacing wires before weaning from bypass during port-access heart surgery can be difficult or impossible. Sometimes, it is necessary to pacing the patient to wean from bypass, and it is problematic to exchange the Edwards pulmonary vent (EndoVent) for a pace catheter under the drapes. Our objective was to devise an effective means of pacing the patient using the pulmonary vent catheter. METHODS: All patients having aortic valve minimally invasive port-access surgery have a pacing wire deployed through the Edwards EndoVent catheter. We did a retrospective chart analysis of these cases. RESULTS: After reviewing the anesthesia records, we determined that we were able to reliably convert the pulmonary vent catheter, which is beneficial for the surgery, into a pacing catheter before weaning from bypass 100% of the time. The mean pacing threshold current was 1.60 mA with the wire in the right ventricular apex. CONCLUSIONS: We found that in all 25 patients we were able to rapidly convert the vent catheter into a reliable pacing catheter without any complications.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA