RESUMO
BACKGROUND: The PRAETORIAN score estimates the risk of failure of subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy by using generator and lead positioning on bidirectional chest radiographs. The PRospective randomized compArative trial of subcutanEous implanTable cardiOverter-defibrillatoR ImplANtation with and without DeFibrillation Testing (PRAETORIAN-DFT) investigates whether PRAETORIAN score calculation is noninferior to defibrillation testing (DFT) with regard to first shock efficacy in spontaneous events. OBJECTIVE: This prespecified subanalysis assessed the predictive value of the PRAETORIAN score for defibrillation success in induced ventricular arrhythmias. METHODS: This multicenter investigator-initiated trial randomized 965 patients between DFT and PRAETORIAN score calculation after de novo S-ICD implantation. Successful DFT was defined as conversion of induced ventricular arrhythmia in <5 seconds from shock delivery within 2 attempts. Bidirectional chest radiographs were obtained after implantation. The predictive value of the PRAETORIAN score for DFT success was calculated for patients in the DFT arm. RESULTS: In total, 482 patients were randomized to undergo DFT. Of these patients, 457 (95%) underwent DFT according to protocol, of whom 445 (97%) had successful DFT and 12 (3%) had failed DFT. A PRAETORIAN score of ≥90 had a positive predictive value of 25% for failed DFT, and a PRAETORIAN score of <90 had a negative predictive value of 99% for successful DFT. A PRAETORIAN score of ≥90 was the strongest independent predictor for failed DFT (odds ratio 33.77; confidence interval 6.13-279.95; P < .001). CONCLUSION: A PRAETORIAN score of <90 serves as a reliable indicator for DFT success in patients with S-ICD, and a PRAETORIAN score of ≥90 is a strong predictor for DFT failure.