1.
J Cardiovasc Electrophysiol ; 34(8): 1690-1697, 2023 08.
Artigo em Inglês | MEDLINE | ID: mdl-37417961

RESUMO

INTRODUCTION: Transseptal puncture (TSP) to allow for large delivery sheath left atrial (LA) access remains a challenging aspect of LA appendage closure (LAAC) in patients with prior history of TSP, thick or lipomatous septum, atrial septal aneurysms, or other complex cardiac anatomies. This study investigates the use of the VersaCross large access (VLA) system (Baylis Medical/Boston Scientific) to improve procedural efficiency of LAAC compared to the standard needle workflow. METHODS AND RESULTS: Fifty LAAC procedures using WATCHMAN FLX between November 2021 and September 2022 were retrospectively analyzed comparing the VLA workflow (n = 25) to the standard needle workflow (n = 25). Study primary endpoint was time to procedural efficiency, and secondary endpoints included TSP time, acute LAAC success, fluoroscopy use, device recaptures, and periprocedural complications. Acute LAAC was successfully completed in all cases with no intraprocedural complications. TSP time was faster, but not significant, using the VLA workflow compared to the standard RF needle workflow (2.6 ± 1.1 min vs. 3.0 ± 1.8 min, p = 0.38). Time to WATCHMAN sheath in LA from TSP was 27% faster (1.5 ± 0.8 min vs. 2.1 ± 0.9 min; p = 0.03), and time to WATCHMAN release from TSP was 19% faster (10.5. ± 2.5 min vs. 13.0 ± 3.7 min; p = 0.01) with the VLA workflow. Overall procedure time was 15% faster (30.4 ± 5.1 min vs. 36.0 ± 6.6 min; p = 0.003) using VLA. Fluoroscopy time was 25% lower (4.0 ± 2.2 min vs. 5.5 ± 2.3 min; p = 0.003) and fluoroscopy dose was 60% lower (97.0 ± 91.7 mGy vs. 241.8 ± 240.6 mGy; p = 0.01) and more consistent [F-test, p ˂ 0.0001] using the VLA workflow compared to the needle workflow. CONCLUSION: The VLA system streamlines LAAC procedures, improving LAAC efficiency and reducing fluoroscopy use by allowing for de novo dilation of the septum for large-bore delivery sheaths, and reducing device exchanges and delivery sheath manipulation.


Assuntos
Apêndice Atrial , Fibrilação Atrial , Acidente Vascular Cerebral , Humanos , Apêndice Atrial/diagnóstico por imagem , Apêndice Atrial/cirurgia , Fibrilação Atrial/diagnóstico por imagem , Fibrilação Atrial/cirurgia , Estudos Retrospectivos , Cateterismo Cardíaco , Átrios do Coração , Resultado do Tratamento
2.
Artigo em Inglês | MEDLINE | ID: mdl-36626294

RESUMO

OBJECTIVE: The radiofrequency (RF) needle has been shown to improve transseptal puncture efficiency and safety compared to mechanical needles. This study aimed to investigate the use of VersaCross RF transseptal wire system (Baylis Medical) to improve procedural efficiency of left atrial appendage closure (LAAC) compared to the standard RF needle-based workflow. METHODS: Eighty-one LAAC procedures using WATCHMAN FLX were retrospectively analyzed comparing the standard RF needle-based workflow to a RF wire-based workflow. Study primary endpoint was time to WATCHMAN device release, and secondary endpoints were transseptal puncture time, LAAC success, fluoroscopy use, and procedural complications. RESULTS: Twenty-five cases using standard RF needle-based workflow were compared to 56 cases using the RF wire-based workflow. Baseline patient characteristics were similar between both groups. LAAC was successful in all patients with no differences in intraprocedural complication rates (p = 0.40). Transseptal puncture time was 1.3 min faster using the RF wire-based workflow compared to the standard RF needle-based workflow (6.5 ± 2.3  vs. 7.8 ± 2.3 min, p = 0.02). Overall, time to final WATCHMAN device release was 4.5 min faster with the RF wire-based workflow compared to the RF needle-based workflow (24.6 ± 5.6 vs. 29.1 ± 9.6 min, p = 0.01). Fluoroscopy time was 21% lower using the RF wire-based workflow (7.6 ± 2.8 vs. 9.6 ± 4.4 min; p = 0.05) and fluoroscopy dose was 67% lower (47.1 ± 35.3 vs. 144.9 ± 156.9 mGy, p = 0.04) and more consistent (F-test, p ˂ 0.0001). CONCLUSIONS: The RF wire-based workflow streamlines LAAC procedures, improving LAAC efficiency and safety by reducing fluoroscopy, device exchanges, and delivery sheath manipulation.

3.
J Cardiovasc Electrophysiol ; 33(8): 1927-1931, 2022 08.
Artigo em Inglês | MEDLINE | ID: mdl-35229417

RESUMO

Transseptal left atrial catheterization is routinely used for many common catheter-based interventions. Tools for transseptal catheterization have advanced over the recent years. Such tools include imaging advances with intracardiac echocardiology as well as an array of needles, wires, and dilators to achieve transseptal access with greater ease and safety. This study will discuss the contemporary tools for transseptal catheterization and guidance for difficult cases.


Assuntos
Apêndice Atrial , Ablação por Cateter , Cateterismo Cardíaco/efeitos adversos , Cateterismo Cardíaco/métodos , Ablação por Cateter/efeitos adversos , Ablação por Cateter/métodos , Átrios do Coração , Septos Cardíacos/diagnóstico por imagem , Septos Cardíacos/cirurgia , Humanos , Punções
4.
Catheter Cardiovasc Interv ; 97(5): 865-868, 2021 04 01.
Artigo em Inglês | MEDLINE | ID: mdl-33539042

RESUMO

We report a modified technique of wire atrial septostomy (WAS) with a reverse transseptal puncture (TSP) in an infant case of pulmonary atresia with intact ventricular septum. A radiofrequency (RF) wire was advanced to the septum through a 4 Fr pigtail catheter hooked on the left side of atrial septum and RF energy was applied while advancing the wire across the septum. Following that reverse TSP, WAS was performed to cut the septal tissue using a 0.010 microwire and RF wire. The atrial septum defect (ASD) was enlarged to a size of 15 mm. WAS with a reverse TSP could be a useful and safe method to enlarge ASD in infants with congenital heart diseases.


Assuntos
Septo Interatrial , Ablação por Cateter , Comunicação Interatrial , Septo Interatrial/diagnóstico por imagem , Septo Interatrial/cirurgia , Comunicação Interatrial/diagnóstico por imagem , Comunicação Interatrial/cirurgia , Humanos , Lactente , Punções , Resultado do Tratamento
5.
J Vasc Surg ; 69(1): 249-252, 2019 01.
Artigo em Inglês | MEDLINE | ID: mdl-30177409

RESUMO

Chronic type B aortic dissection with aneurysmal degeneration requiring intervention presents significant therapeutic challenges. Thoracic endovascular aortic repair with a fenestrated endograft is a feasible option, but false lumen branches without an adjacent re-entry or perforation in the septum can pose a significant challenge. We present two cases of fenestrated endovascular aneurysm repair for chronic type B aortic dissection in which a renal artery from the false lumen was cannulated by creating a "neofenestration" through the dissection flap using a radiofrequency PowerWire (Baylis Medical Inc, Montreal, Quebec, Canada) technique (Toronto PowerWire fenestration technique).


Assuntos
Aneurisma da Aorta Torácica/cirurgia , Dissecção Aórtica/cirurgia , Implante de Prótese Vascular/métodos , Procedimentos Endovasculares/métodos , Idoso , Dissecção Aórtica/diagnóstico por imagem , Aneurisma da Aorta Torácica/diagnóstico por imagem , Aortografia/métodos , Prótese Vascular , Implante de Prótese Vascular/instrumentação , Doença Crônica , Angiografia por Tomografia Computadorizada , Procedimentos Endovasculares/instrumentação , Feminino , Humanos , Masculino , Radiografia Intervencionista , Stents , Resultado do Tratamento , Dispositivos de Acesso Vascular
6.
Am J Kidney Dis ; 69(2): 309-313, 2017 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-27866966

RESUMO

Stenosis or occlusion of central veins in hemodialysis patients is common, especially with previous intravascular catheter or device use. Superior vena cava (SVC) obstruction is emerging as a frequent chronic complication of central vein cannulation that not only jeopardizes the availability of vascular access for hemodialysis, but can become a life-threatening emergency. Clinical features of SVC syndrome can be subtle or dramatic, including facial swelling and shortness of breath, which require expeditious attention and intervention. The approach to SVC syndrome involves judicious use of imaging techniques to define the cause and location. Early management with endovascular intervention with angioplasty and stent placement is the usual first choice. The occlusion can often be recanalized using new techniques such as radiofrequency wire and then salvaged with stents, providing prompt resolution of symptoms. Limitations to interventions include requirement of cutting-edge equipment, expertise, expense, and the usually temporary nature of the resolution. Surgery is considered the treatment of last resort for refractory cases. SVC syndrome can be prevented by minimizing catheter and intravascular device use through early recognition of patients with chronic kidney disease, early referral for education about all choices for kidney replacement modalities, and early placement of arteriovenous access prior to the onset of dialysis therapy.


Assuntos
Cateterismo Venoso Central/efeitos adversos , Diálise Renal , Síndrome da Veia Cava Superior/etiologia , Humanos , Masculino , Pessoa de Meia-Idade
7.
J Interv Card Electrophysiol ; 64(1): 183-190, 2022 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-35194727

RESUMO

PURPOSE: Conventional catheter ablation for atrial fibrillation requires fluoroscopy, which has inherent risks of radiation exposure to patients and medical staff. Optimization of fluoroscopy parameters and use of three-dimensional electroanatomic mapping (EAM) and intracardiac echocardiography (ICE) have helped to reduce radiation exposure; however, despite growing evidence, there are still concerns about safety and added procedure time associated with fluoroless procedures, particularly in left-sided ablations, due to the potential risk of complications. Herein, we report our initial experience using a radiofrequency (RF) wire for completely fluoroless radiofrequency ablation (RFA) and cryoballoon ablation (CBA). METHODS: A retrospective analysis was conducted on ablation procedures for various cardiac arrhythmias performed non-fluoroscopically at two centers using the VersaCross RF wire transseptal system under EAM and ICE guidance. RESULTS: A total of 72 and 54 patients underwent RFA and CBA, respectively, successfully without any procedural complications. Transseptal access time for RFA was 14.5 ± 6.6 min from procedure start (including sheath and catheter placements ± right-sided ablation) or 2.8 ± 1.0 min from RF wire insertion into the femoral introducer. Transseptal access time for CBA was 19.2 ± 11.7 min from procedure start (including sheath and catheter placements ± right-sided ablation) or 3.5 ± 1.6 min from RF wire insertion into the femoral introducer. Average procedure time was 104.4 ± 38.0 min for RFA and 91.1 ± 22.1 min for CBA. CONCLUSIONS: A RF wire can be used to achieve completely fluoroless transseptal puncture safely and effectively while improving procedural efficiency in both RFA and CBA.


Assuntos
Fibrilação Atrial , Ablação por Cateter , Fibrilação Atrial/diagnóstico por imagem , Fibrilação Atrial/cirurgia , Ablação por Cateter/métodos , Fluoroscopia , Humanos , Punções , Estudos Retrospectivos , Resultado do Tratamento
8.
J Vasc Surg Cases Innov Tech ; 5(3): 356-359, 2019 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-31440713

RESUMO

The goal of this case report was to demonstrate the feasibility, safety, and efficacy of image fusion software to treat symptomatic central venous occlusion with radiofrequency wire after failure of conventional endovascular techniques. The complex and chronic central venous occlusion was successfully treated without complications. The combination of these techniques provided an endovascular solution and could be considered in select cases.

9.
J Vasc Surg Cases Innov Tech ; 5(4): 553-556, 2019 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-31799483

RESUMO

We report a case of a frozen elephant trunk arch repair, where the stent graft was unintentionally placed into the false lumen. Postoperative imaging demonstrated an enlarged false lumen with no thoracic aorta fenestrations that could be traversed to place another thoracic endovascular aortic repair endograft into the true lumen. An atraumatic radiofrequency wire (PowerWire, Baylis Medical, Montreal, Quebec, Canada) was used to create a new septal fenestration, enabling thoracic endovascular aortic repair endograft extension into the thoracic true lumen. This novel use of a radiofrequency wire can enable safe and controlled endovascular septal fenestration even in chronic dissections to redirect flow into the true lumen.

10.
Cardiovasc Intervent Radiol ; 42(1): 130-136, 2019 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-30288588

RESUMO

INTRODUCTION: Chronically occluded venous stents may be difficult to salvage, necessitating multiple approaches and adjunctive techniques. Radiofrequency wires have been used to cross chronic obstructions in a variety of settings. Herein, radiofrequency wire recanalization (RFWR) of chronically occluded venous stents is presented. MATERIALS AND METHODS: Fifteen patients (8 males; 7 females) aged 23-74 years (median 42 years) underwent prior venous stenting in the setting of venous occlusions. Seven had an underlying coagulopathy, three had May-Thurner syndrome, three had provoked deep venous thrombosis, one had a dialysis fistula, and one had pacer leads in situ. All indwelling venous stents were chronically occluded and symptomatic. Out of 15 patients, 13 (87%) had iliocaval venous stents, while two (13%) had superior vena cava and/or upper extremity venous stents. After failing conventional and blunt recanalization techniques, RFWR was performed to traverse the chronic occlusions. Technical success and complications were recorded. RESULTS: Fifteen patients underwent 19 procedures. Twelve patients underwent one procedure, two patients had two procedures, and one patient had three procedures. RFWR alone was used in 12/19 (63%) procedures, while 7/19 (37%) required adjunctive sharp recanalization techniques. Technical success or crossing of the occlusion with flow restoration through the occluded segments was achieved in 17/19 (89%). One major and one minor complication occurred. CONCLUSION: Radiofrequency wire recanalization is an effective adjunct to revise chronically occluded venous stents, potentially increasing procedural success in challenging cases.


Assuntos
Oclusão de Enxerto Vascular/cirurgia , Ablação por Radiofrequência/instrumentação , Ablação por Radiofrequência/métodos , Stents , Adulto , Idoso , Desenho de Equipamento , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Flebografia/métodos , Estudos Retrospectivos , Resultado do Tratamento , Adulto Jovem
11.
CVIR Endovasc ; 1(1): 24, 2018.
Artigo em Inglês | MEDLINE | ID: mdl-30652155

RESUMO

BACKGROUND: Radiofrequency (RF) wire recanalization of short segments of central venous obstruction has been considered safe; however its use for recanalization of long segments of inferior vena cava (IVC) has not been reported. CASE PRESENTATION: A 55-year-old female with recurrent massive hematemesis was found to have systemic venous upper esophageal varices on endoscopy and an extensive chronic IVC occlusion on CT. Using both a percutaneous transhepatic and transfemoral approach IVC recanalization was performed. A snare was advanced to the cavo-atrial junction via transhepatic venous access. From the groin utilizing RF wire steerable guide sheaths, endovascular reconstruction of the IVC was performed. Post recanalization venography demonstrated patent stented IVC and marked decrease in the intraabdominal-pelvic collaterals. No recurrence of hematemesis was noted. After 6 months, patient remained asymptomatic and had functioning right femoral arteriovenous hemodialysis graft. CONCLUSIONS: Using appropriate techniques, Power wire recanalization of long occlusive segments of IVC can be safe and effective.

12.
Cardiovasc Intervent Radiol ; 39(7): 1040-4, 2016 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-26902703

RESUMO

Radiofrequency (RF) guide wires have been applied to cardiac interventions, recanalization of central venous thromboses, and to cross biliary occlusions. Herein, the use of a RF wire technique to revise chronically occluded transjugular intrahepatic portosystemic shunts (TIPS) is described. In both cases, conventional TIPS revision techniques failed to revise the chronically thrombosed TIPS. RF wire recanalization was successfully performed through each of the chronically thrombosed TIPS, demonstrating initial safety and feasibility in this application.


Assuntos
Ablação por Cateter/instrumentação , Ablação por Cateter/métodos , Cateterismo/instrumentação , Cateterismo/métodos , Derivação Portossistêmica Transjugular Intra-Hepática , Complicações Pós-Operatórias/cirurgia , Trombose Venosa Profunda de Membros Superiores/cirurgia , Angiografia Digital , Desenho de Equipamento , Humanos , Masculino , Pessoa de Meia-Idade , Resultado do Tratamento