Your browser doesn't support javascript.
loading
Evaluation of an automatized AMBU ESSI-1 INC designed at the Instituto Nacional de Cardiología Ignacio Chávez / Evaluación del AMBU automatizado ESSI-1 INC diseñado en el Instituto Nacional de Cardiología Ignacio Chávez
Abundes-Velasco, Arturo; Sánchez-Pérez, Efraín; Jiménez-Rodríguez, Gian M.; Molina-Méndez, Francisco J.; Herrera-Elizalde, Rafael; Pulido-Zamudio, Tomás; Álvarez-Rosales, Humberto; Villegas-Juache, Juan; Aceves-Díaz-González, Sebastián; Parra-Rodríguez, David A.; Cruz-Enríquez, Francisco J.; Cobos-Reyes, Farid A.; Aguilar-Torres, Eduardo A.; Islas-Navarro, Enid A.; Nuñez-Gómez, José C.; Peña-Duque, Marco A..
Afiliação
  • Abundes-Velasco, Arturo; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Sánchez-Pérez, Efraín; Instituto Nacional de Cardiología Ignacio Chávez. Area of Cardiovascular Technological Innovation. MX
  • Jiménez-Rodríguez, Gian M.; Instituto Nacional de Cardiología Ignacio Chávez. Cardiovascular Intensive Post-Surgical Therapy. MX
  • Molina-Méndez, Francisco J.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Anesthesiology. MX
  • Herrera-Elizalde, Rafael; Instituto Nacional de Cardiología Ignacio Chávez. Department of Anesthesiology. MX
  • Pulido-Zamudio, Tomás; Instituto Nacional de Cardiología Ignacio Chávez. Department of Cardioneumology. MX
  • Álvarez-Rosales, Humberto; Instituto Nacional de Cardiología Ignacio Chávez. Department of Anesthesiology. MX
  • Villegas-Juache, Juan; Instituto Nacional de Cardiología Ignacio Chávez. Bioterium. MX
  • Aceves-Díaz-González, Sebastián; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Parra-Rodríguez, David A.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Cruz-Enríquez, Francisco J.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Cobos-Reyes, Farid A.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Aguilar-Torres, Eduardo A.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Islas-Navarro, Enid A.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Nuñez-Gómez, José C.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
  • Peña-Duque, Marco A.; Instituto Nacional de Cardiología Ignacio Chávez. Department of Innovation and Technology. MX
Arch. cardiol. Méx ; 93(4): 391-397, Oct.-Dec. 2023. tab, graf
Article em En | LILACS-Express | LILACS | ID: biblio-1527715
Biblioteca responsável: MX1.1
ABSTRACT
Abstract

Objective:

The objectives of this study were to evaluate an automated device for ventilatory support based on AMBU manufactured in March 2020.

Methods:

The ESSI-1 INC was evaluated through pulmonary mechanics and physiology parameters through compensatory spirometer tests (TISSOTs), and an artificial lung Model5600i Dual Adult PNEU VIEW SYSTEM; it was also compared to the anesthetic ventilatory support equipment (AEONMED 7500) in porcine models, measuring ventilatory, hemodynamic and gasometric parameters.

Results:

This equipment (ESSI-1 INC) was successfully tested by mechanical and biological models, such as pigs in which its performance was evaluated in terms of variability of tidal volume, ventilation frequency, and I/E relationship versus the manual performance of two medical interns. All the results turned out as expected and were satisfactory.

Conclusions:

It is safe and effective equipment and should be tested and used in diverse clinical conditions to standardize the ventilatory safety and care of patients who require it.
RESUMEN
Resumen

Objetivo:

Evaluar un dispositivo automatizado para la asistencia ventilatoria basado en un AMBU manufacturado en Marzo del 2020.

Métodos:

El ESSI-1 INC fue evaluado por medio de parámetros fisiológicos y mecánica pulmonar a través de pruebas de espirómetro compensatorios (TISSOT); pulmón artificial (Modelo 5600i Dual Adult PNEU VIEW SYSTEM); así como su desempeño comparado a la máquina de anestesia (AEONMED 7500) en modelos porcinos, midiendo criterios ventilatorios, hemodinámicos y gasométricos.

Resultados:

Este equipo (ESSI-1 INC) fue exitosamente probado por modelos mecánicos y biológicos, tales como cerdos donde su desempeño fue evaluado en términos de la variabilidad del volumen tidal, frecuencia ventilatoria, y relación I/E versus el desempeño manual de dos médicos. Todos los resultados finalizaron como se esperaba de forma satisfactoria.

Conclusiones:

Es un equipo seguro y efectivo, el cual debería ser probado y usado en distintas condiciones clínicas para estandarizar la seguridad ventilatoria y cuidado de pacientes que lo requieran.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Tipo de estudo: Guideline / Prognostic_studies Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Tipo de estudo: Guideline / Prognostic_studies Idioma: En Ano de publicação: 2023 Tipo de documento: Article