BACKGROUND: Although ventricular assist devices (VAD) have improved survival in selected patients, their use continues to be complicated by thromboembolism and end-organ failure. Complement activation may play a role in the pathogenesis of these complications. Previous studies have found that the complement common terminal pathway is activated during VAD circulation. C3a levels rise dramatically during VAD use. Because the C3a fragment is generated by either the alternative or classical pathway, the purpose of this study is to determine the relative importance of the respective pathways in complement activation during in vitro VAD circulation. METHODS: Six in vitro VAD circuits were simulated for 3 days using 450 mL of human blood. Temperature, activated clotting time, pH, pCO2, pO2, Ca2+, and glucose were maintained at physiologic levels. Enzyme immunoassays were used to measure concentrations of fragment Bb to indicate alternative pathway activation and fragment C4d to indicate classical pathway activation. RESULTS: Fragment Bb concentrations rise from 1.92 to 10.77 micrograms/mL during the first 6 hours of circulation. Thereafter, Bb levels plateau. C4d concentrations slowly rise from a baseline of 1.49 to 6.84 micrograms/mL in 72 hours. CONCLUSIONS: These findings suggest that both the alternative and classical pathways of complement are activated during VAD circulation. Alternative pathway activation precedes classical pathway activation during in vitro VAD circulation and may be of greater clinical importance during clinical VAD circulation.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Complemento C4b
/
Coração Auxiliar
/
Ativação do Complemento
/
Via Alternativa do Complemento
/
Insuficiência Cardíaca
Limite:
Humans
Idioma:
En
Ano de publicação:
1999
Tipo de documento:
Article