Your browser doesn't support javascript.
loading
Prospective multiaxial motion correction for fMRI.
Ward, H A; Riederer, S J; Grimm, R C; Ehman, R L; Felmlee, J P; Jack, C R.
Afiliação
  • Ward HA; Department of Radiology, Mayo Clinic, Rochester, Minnesota 55905, USA.
Magn Reson Med ; 43(3): 459-69, 2000 Mar.
Article em En | MEDLINE | ID: mdl-10725890
ABSTRACT
Corruption of the image time series due to interimage head motion limits the clinical utility of functional MRI. This paper presents a method for real-time prospective correction of rotation and translation in all six degrees of rigid body motion. By incorporating an orbital navigator (ONAV) echo for each of the sagittal, axial, and coronal planes into the fMRI pulse sequence, rotation and translation can be measured and the spatial orientation of the image acquisition sequence that follows can be corrected prospectively in as little as 160 msec. Testing of the method using a computerized motion phantom capable of performing complex multiaxial motion showed subdegree rotational and submillimeter translational accuracy over a range of +/-8 degrees and +/-8 mm of motion. In vivo images demonstrate correction of simultaneous through-plane and in-plane motion and improved detection of fMRI activation in the presence of head motion.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Aumento da Imagem / Movimentos da Cabeça Limite: Humans Idioma: En Ano de publicação: 2000 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Aumento da Imagem / Movimentos da Cabeça Limite: Humans Idioma: En Ano de publicação: 2000 Tipo de documento: Article