A micromachined silicon depth probe for multichannel neural recording.
Yoon, T H; Hwang, E J; Shin, D Y; Park, S I; Oh, S J; Jung, S C; Shin, H C; Kim, S J.
Afiliação
  • Yoon TH; Inter-university Semiconductor Research Center, Seoul National University, Korea.
IEEE Trans Biomed Eng ; 47(8): 1082-7, 2000 Aug.
Article em En | MEDLINE | ID: mdl-10943057
ABSTRACT
A process of making a new type of silicon depth-probe microelectrode array is described using a combination of plasma and wet etch. The plasma etch, which is done using a low temperature oxide (LTO) mask, enables probe thickness to be controlled over a range from 5 to 90 mu. Bending tests show that the probe's mechanical strength depends largely on shank thickness. More force can be applied to thicker shanks while thinner shanks are more flexible. One can then choose a thickness and corresponding mechanical strength using the process developed. The entire probe shaping process is performed only at low temperature, and thus is consistent with the standard CMOS fabrication. Using the probe in recording from rat's somatosensory cortex, we obtained four channel simultaneous recordings which showed clear independence among channels with a signal-to-noise ratio performance comparable with that obtained using other devices.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Microeletrodos / Neurônios Limite: Animals Idioma: En Ano de publicação: 2000 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Microeletrodos / Neurônios Limite: Animals Idioma: En Ano de publicação: 2000 Tipo de documento: Article