Cardiopulmonary monitoring at home: the CHIME monitor.
Physiol Meas
; 22(2): 267-86, 2001 May.
Article
em En
| MEDLINE
| ID: mdl-11411239
ABSTRACT
A new physiologic monitor for use in the home has been developed and used for the Collaborative Home Infant Monitor Evaluation (CHIME). This monitor measures infant breathing by respiratory inductance plethysmography and transthoracic impedance; infant electrocardiogram, heart rate and R-R interval; haemoglobin O2 saturation of arterial blood at the periphery and sleep position. Monitor signals from a representative sample of 24 subjects from the CHIME database were of sufficient quality to be clinically interpreted 91.7% of the time for the respiratory inductance plethysmograph, 100% for the ECG, 99.7% for the heart rate and 87% for the 16 subjects of the 24 who used the pulse oximeter. The monitor detected breaths with a sensitivity of 96% and a specificity of 65% compared to human scorers. It detected all clinically significant bradycardias but identified an additional 737 events where a human scorer did not detect bradycardia. The monitor was considered to be superior to conventional monitors and, therefore, suitable for the successful conduct of the CHIME study.