Your browser doesn't support javascript.
loading
Measurements of electron beam damage for organic crystals in a high voltage electron microscope with image plates.
Ohno, Tamotsu; Sengoku, Masaya; Arii, Tatsuo.
Afiliação
  • Ohno T; Department of Physics, Aichi Medical University, Nagakute-cho, Aichi-ken, Japan. ohno@aichi-med-u.ac.jp
Micron ; 33(4): 403-6, 2002.
Article em En | MEDLINE | ID: mdl-11814879
ABSTRACT
The critical dose for extinction of the diffraction pattern from behenic acid monolayer crystals increased with increasing accelerating voltages. The mean values at 114, 500 and 1000kV were 0.6, 1.8 and 2.2 electrons/A(2), respectively. The critical dose at 500-1000kV is three or more times as large as that at 114kV. Considering this with the recent measured value of the detective quantum efficiency of the image plate, 0.5 at 500kV and 0.4 at 1000kV, 1.5 times as much information can be collected from a crystal at 500-1000kV as at 114kV. Therefore, the combined use of high voltage electron microscopy and image plate detectors offers a significant improvement over conventional electron microscopy approaches for the study of radiation sensitive organic molecule crystals.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Química Orgânica Idioma: En Ano de publicação: 2002 Tipo de documento: Article