Your browser doesn't support javascript.
loading
Biosensors and flow-through system for the determination of creatinine in hemodialysate.
Erlenkötter, Ansgar; Fobker, Manfred; Chemnitius, Gabriele-Christine.
Afiliação
  • Erlenkötter A; Institute for Chemical and Biochemical Sensor Research (ICB), Münster, Germany.
Anal Bioanal Chem ; 372(2): 284-92, 2002 Jan.
Article em En | MEDLINE | ID: mdl-11936100
ABSTRACT
Biosensors for the determination of creatinine have been developed and integrated into a flow-through system. The sensors are based on a screen-printed three electrode transducer with a platinum working electrode. Applying the multi-enzyme sequence of creatininase (CA), creatinase (CI) and sarcosine oxidase (SO) hydrogen peroxide has been detected amperometrically. An optimal enzyme load was found to be 4.4 U/0.28 U/0.20 U (CA/CI/SO) and 0.28 U/0.20 U (CI/SO) per electrode for the creatinine sensor and for the creatine sensor, respectively. Among a variety of polymers Nafion has shown the highest efficiency to exclude interfering substances like ascorbic acid, acetaminophen and uric acid. First determinations of creatinine in dialysate samples obtained during hemodialysis treatments have shown a good correlation to the conventional methods, the Jaffé reaction (y=0.945x+ 2.8, R=0.9882, n=9) and the enzymatic photometric method (y=0.891x+3.5, R=0.9917, n=9).
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Soluções para Hemodiálise / Creatinina Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2002 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas Biossensoriais / Soluções para Hemodiálise / Creatinina Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2002 Tipo de documento: Article