Coronary perforation as a potential complication derived from coronary thrombectomy with the X-Sizer device.
Sanmartín, Marcelo; Goicolea, Javier; Ruiz-Salmerón, Rafael; Mantilla, Ramón; Sterling, Juán; Meneses, Daniel; Calvo, Francisco.
Afiliação
  • Sanmartín M; Interventional Cardiology Department, Medtec, Hospital Meixoeiro, Vigo, Pontevedra, Spain. javier.goicolea.ruigomez@sergas.es
Catheter Cardiovasc Interv ; 56(3): 378-82, 2002 Jul.
Article em En | MEDLINE | ID: mdl-12112893
ABSTRACT
We report a case of coronary perforation after failed atherectomy with a 2 mm X-Sizer catheter in recent totally occluded right coronary artery. The perforation was successfully managed with a polytetrafluoroethylene-covered stent with satisfactory final angiographic results. Possible predictors of this complication with this new device are discussed.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombose Coronária / Cateterismo / Trombectomia / Vasos Coronários / Complicações Intraoperatórias Tipo de estudo: Prognostic_studies Limite: Humans / Male / Middle aged Idioma: En Ano de publicação: 2002 Tipo de documento: Article