The sensitive determination of abanoquil in blood by high-performance liquid chromatography/atmospheric pressure ionization mass spectrometry.
Kaye, B; Clark, M W; Cussans, N J; Macrae, P V; Stopher, D A.
Afiliação
  • Kaye B; Department of Drug Metabolism Pfizer Central Research, Sandwich, Kent, UK.
Biol Mass Spectrom ; 21(11): 585-9, 1992 Nov.
Article em En | MEDLINE | ID: mdl-1360817
ABSTRACT
A method is described for the determination of abanoquil in human blood. The method is based on high-performance liquid chromatography (HPLC)/atmospheric pressure positive ion chemical ionization mass spectrometry, using (2H3)abanoquil as internal standard. Multiple reaction monitoring is employed for selectivity and sensitivity, which enables quantification over the range 10-500 pg ml-1 with acceptable precision and accuracy. This assay methodology illustrates the versatility of atmospheric pressure ionization/tandem mass spectrometry, in conjunction with HPLC, for the separation and quantification of drugs in the subnanogram per millilitre range.