Your browser doesn't support javascript.
loading
Pitfalls of bone mineral density evaluation by dual-photon absorptiometry.
Elliott, M W; Eisenberg, B; Dorin, M H; Vela, S; Dorin, R I; Murata, G H; Tzamaloukas, A H.
Afiliação
  • Elliott MW; Department of Radiology, Albuquerque Veterans Affairs Medical Center, New Mexico.
Clin Nucl Med ; 17(9): 714-20, 1992 Sep.
Article em En | MEDLINE | ID: mdl-1395343
ABSTRACT
Material absorbing photons aligned with the lumbar vertebrae can create falsely elevated measurements of bone mineral density during dual-photon absorptiometry. Three cases illustrating this phenomenon are presented. Although bone mineral density was overestimated in each case, calculated fracture risk was normal in two cases and greatly increased in the third. Photon-absorbing material can create overestimates of bone mineral density during dual-photon absorptiometry, even when a greatly increased fracture risk is computed.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Absorciometria de Fóton / Densidade Óssea / Vértebras Lombares Limite: Adult / Aged / Female / Humans / Male / Middle aged / Pregnancy Idioma: En Ano de publicação: 1992 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Absorciometria de Fóton / Densidade Óssea / Vértebras Lombares Limite: Adult / Aged / Female / Humans / Male / Middle aged / Pregnancy Idioma: En Ano de publicação: 1992 Tipo de documento: Article