The behavior of a carbon paste electrode was investigated as an amperometric detector for the determination of dipyrone by flow injection analysis (FIA). The electrode presented low cost and easy construction by simple mixing of graphite powder and mineral oil. Initially, an electrochemical study of the dipyrone oxidation at a carbon paste electrode has been developed before its use in the FIA system. The oxidation currents monitored at +0.35 V versus Ag/AgCl, were proportional to the dipyrone concentrations. Experimental parameters, such as nature of supporting electrolyte, pH of the carrier solution, flow rate, sample volume injection and probable interferents were investigated. Under the best experimental conditions selected, the calibration curve for dipyrone was linear in the concentration range from 4.91 x 10(-6) to 2.50 x 10(-4) M l(-1) (I(anodic)/microA)=0.056+81.06 [dipyrone]) with a detection limit of 2.07 x 10(-6) M l(-1). Recoveries ranged from 93.8 to 100.8% and an analytical frequency of 130 h(-1) was achieved. The proposed flow procedure has been satisfactorily applied to the determination of dipyrone in several pharmaceutical formulations.