Your browser doesn't support javascript.
loading
Analysis of iodinated X-ray contrast agents in water samples by ion chromatography and inductively-coupled plasma mass spectrometry.
Sacher, Frank; Raue, Brigitte; Brauch, Heinz-Jürgen.
Afiliação
  • Sacher F; DVGW-Technologiezentrum Wasser, Karlsruher Strasse 84, 76139 Karlsruhe, Germany. sacher@tzw.de
J Chromatogr A ; 1085(1): 117-23, 2005 Aug 26.
Article em En | MEDLINE | ID: mdl-16106857
ABSTRACT
In this paper, an analytical method for the determination of six iodinated X-ray contrast agents (amidotrizoic acid, iohexol, iomeprol, iopamidol, iopromide, and ioxitalamic acid), iodide, and iodate in water samples is presented. The method is based on a separation of the analytes by ion chromatography (IC) and a subsequent detection by inductively-coupled plasma mass spectrometry (ICP-MS). The method was optimised with respect to separation conditions (column type and eluent composition) and extensively validated. Without pre-concentration of the samples, limits of detection below 0.2 microg/l could be achieved whereby reproducibility was below 6% for all compounds under investigation.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Poluentes Químicos da Água / Cromatografia por Troca Iônica / Meios de Contraste Idioma: En Ano de publicação: 2005 Tipo de documento: Article
Buscar no Google