[Validation of PM10 and PM2.5 personal samplers: comparison between PEM and CEN-ARPAV selectors]. / Studio di validazione dei selettori per il campionamento personale di PM10 e PM2.5: confronto fra selettori PEM e teste di prelievo CEN-ARPAV.
G Ital Med Lav Ergon
; 27(3): 362-6, 2005.
Article
em It
| MEDLINE
| ID: mdl-16240596
In order to validate specific personal selectors for the collection of fine particles, 65 double measurements of PM10 and PM2.5 were carried out at the same collection site; thus allowing a comparison between two different methods of sampling. The first method was that normally used by ARPAV in order to sample the above-mentioned granulometric fractions in outdoor environment. The second method was chosen by us for personal sampling (PEM working at 2 l/min and at 4 l/min). In both cases the filters gravimetric analysis conformed to the expectations of D.M. 60 of 2/4/2002. The comparison between the two methods showed a good correlation in both the granulometric fractions: correlation coefficients r for the PM2.5 are equal to 0.96 and 0.99 for the PEM working at 2 l/min and at 4 l/min respectively; r for the PM10 are equal to 0.98 and 0.99 for the PEM working at 2 l/min and at 4 l/min respectively. The analysis of results in terms of fine particles concentration and difference between methods against their mean, shows a slight overestimate of the particles concentration with PEM working at 2 l/min, compared to those working at 4 l/min. Nevertheless, considering the good results obtained even with a flow of 2 l/min, we believe that using PEM working at 2 l/min to monitoring 24 hours-personal exposure assures an improved capacity in the battery-operated pumps.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Monitoramento Ambiental
/
Poluentes Atmosféricos
/
Exposição Ambiental
Tipo de estudo:
Evaluation_studies
Limite:
Humans
Idioma:
It
Ano de publicação:
2005
Tipo de documento:
Article