Technical report: Magnetic resonance direct thrombus imaging at 3 T field strength in patients with lower limb deep vein thrombosis: a feasibility study.
Schmitz, S A; O'Regan, D P; Gibson, D; Cunningham, C; Fitzpatrick, J; Allsop, J; Larkman, D J; Hajnal, J V.
Afiliação
  • Schmitz SA; Imaging Sciences Department, Imperial College, Hammersmith Hospital, London, UK.
Clin Radiol ; 61(3): 282-6, 2006 Mar.
Article em En | MEDLINE | ID: mdl-16488211
ABSTRACT

AIM:

To investigate the feasibility of imaging lower limb deep vein thrombosis using magnetic resonance imaging (MRI) at 3.0 T magnetic field strength with an optimized a T1 magnetization prepared rapid gradient echo technique (MP-RAGE) in patients with normal volunteers as controls. MATERIALS AND

METHODS:

Patients with deep vein thrombosis (n = 4), thrombophlebitis (n = 2) and healthy volunteers (n = 9) were studied. MRI of the distal thigh and upper calf was performed at 3.0 T with MP-RAGE using two pre-pulses to suppress blood and fat (flip angle 15 degrees, echo time 5 ms, and repetition time 10 ms). A qualitative analysis was performed for detection of thrombi and image quality. Contrast-to-noise ratios were determined in thrombosed and patent veins.

RESULTS:

Thrombi were clearly visible as high-signal intensity structures with good suppression of the anatomical background. A blinded reader accurately diagnosed 15 out of 16 cases. The contrast-to-noise ratio measurements showed a positive contrast of thrombus over background muscle 16.9 (SD 4.3, 95% CI 12.5-21.3) and a negative contrast of the lumen to muscle in patent veins of normal volunteers -7.8 (SD 4.3, 95% CI -11.1 to -4.5), with p = 0.0015.

CONCLUSION:

Thrombi generate high signal intensity at 3.0 T allowing for their direct visualization if flowing blood, stationary blood and fat are sufficiently suppressed. This preliminary data supports the development of these techniques for other vascular applications.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Trombose Venosa Tipo de estudo: Qualitative_research Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google