Your browser doesn't support javascript.
loading
The value of multichannel MEG and EEG in the presurgical evaluation of 70 epilepsy patients.
Knake, S; Halgren, E; Shiraishi, H; Hara, K; Hamer, H M; Grant, P E; Carr, V A; Foxe, D; Camposano, S; Busa, E; Witzel, T; Hämäläinen, M S; Ahlfors, S P; Bromfield, E B; Black, P M; Bourgeois, B F; Cole, A J; Cosgrove, G R; Dworetzky, B A; Madsen, J R; Larsson, P G; Schomer, D L; Thiele, E A; Dale, A M; Rosen, B R; Stufflebeam, S M.
Afiliação
  • Knake S; Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA 02129, USA. knake@staff.uni-marburg.de
Epilepsy Res ; 69(1): 80-6, 2006 Apr.
Article em En | MEDLINE | ID: mdl-16516443
ABSTRACT

OBJECTIVE:

To evaluate the sensitivity of a simultaneous whole-head 306-channel magnetoencephalography (MEG)/70-electrode EEG recording to detect interictal epileptiform activity (IED) in a prospective, consecutive cohort of patients with medically refractory epilepsy that were considered candidates for epilepsy surgery.

METHODS:

Seventy patients were prospectively evaluated by simultaneously recorded MEG/EEG. All patients were surgical candidates or were considered for invasive EEG monitoring and had undergone an extensive presurgical evaluation at a tertiary epilepsy center. MEG and EEG raw traces were analysed individually by two independent reviewers.

RESULTS:

MEG data could not be evaluated due to excessive magnetic artefacts in three patients (4%). In the remaining 67 patients, the overall sensitivity to detect IED was 72% (48/67 patients) for MEG and 61% for EEG (41/67 patients) analysing the raw data. In 13% (9/67 patients), MEG-only IED were recorded, whereas in 3% (2/67 patients) EEG-only IED were recorded. The combined sensitivity was 75% (50/67 patients).

CONCLUSION:

Three hundred and six-channel MEG has a similarly high sensitivity to record IED as EEG and appears to be complementary. In one-third of the EEG-negative patients, MEG can be expected to record IED, especially in the case of lateral neocortical epilepsy and/or cortical dysplasia.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cuidados Pré-Operatórios / Magnetoencefalografia / Epilepsias Parciais / Eletroencefalografia Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies Limite: Adolescent / Adult / Child / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cuidados Pré-Operatórios / Magnetoencefalografia / Epilepsias Parciais / Eletroencefalografia Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies Limite: Adolescent / Adult / Child / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2006 Tipo de documento: Article