Your browser doesn't support javascript.
loading
Dural defect repair with fascia by a CO2 laser system in a porcine model.
Forer, Boaz; Vasilyev, Tamar; Brosh, Tamar; Kariv, Naam; Trejo, Leonor Leider; Gil, Ziv; Katzir, Abraham; Fliss, Dan M.
Afiliação
  • Forer B; Department of Otolaryngology Head and Neck Surgery, Tel-Aviv Sourasky Medical Center, Tel Aviv University, Tel Aviv, Israel. forerbr@netvision.net.il
Laryngoscope ; 116(6): 1002-6, 2006 Jun.
Article em En | MEDLINE | ID: mdl-16735918
HYPOTHESIS/OBJECTIVES: We tested the effectiveness of a temperature-controlled CO2 laser soldering system on a porcine model for dural defect reconstruction using a fascial patch. METHODS: A dural patch was excised and then reconstructed with fascia by a CO2 laser system in vitro in 27 animals and in vivo in five animals. RESULTS: After dural reconstruction, the average burst pressure of the soldered patch in vitro, as measured by a custom-made pressure detector, was 258.5 cm H2O. All five pigs in the in vivo group showed no neurological complications or cerebrospinal fluid leak, and the underlying brain tissue showed no thermal injury. CONCLUSION: The CO2 laser system created a watertight bond and did not cause thermal injury to the brain. The procedure is potentially faster than conventional repair, and wound healing may also be more rapid.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dura-Máter / Terapia a Laser / Fasciotomia Limite: Animals Idioma: En Ano de publicação: 2006 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dura-Máter / Terapia a Laser / Fasciotomia Limite: Animals Idioma: En Ano de publicação: 2006 Tipo de documento: Article