Impedance spectroscopy analysis of cell-electrode interface.
Conf Proc IEEE Eng Med Biol Soc
; 2005: 7608-11, 2005.
Article
em En
| MEDLINE
| ID: mdl-17282042
ABSTRACT
Many chronically implanted electrodes suffer sensitivity loss in their applications in brain computer interface systems. It is hard to diagnose the cause of the problem because few measures are available to analyze directly what happened on the cell-electrode interface. In this paper, the impedance characterization of the cell-electrode interface was discussed in detail using equivalent circuit approach, which was used to evaluate the cause of the electrode sensitivity loss. The impedance spectroscopy of the cell-electrode interface acts as a function of several parameters, such as the sealing resistance and the shunt capacitance between the microelectrode and the electrolyte. Changes of the impedance spectroscopy can be traced to the parameter changes of the equivalent circuit, which reflect the status of the cell-electrode interface, such as the cell-electrode gap change, the erosion of microelectrodes, and so on. The circuit impedance simulation results give an important reference for the monitor of the cell-electrode connection, and are also helpful for the improvement of the microelectrode design.