Clinical validation of a new blood collection tube for the accuracy of total homocysteine measurement by different methods.
Bissé, Emmanuel; Epting, Thomas; Kögel, Gabriele; Obeid, Rima; Gempel, Klaus; Huzly, Daniela; Fuhrmann, Klaus; Rolinski, Boris; Herrmann, Wolfgang; Wieland, Heinrich.
Afiliação
  • Bissé E; Department of Clinical Chemistry, University Medical Center Freiburg, Hugstetterstrasse 55, D-79106 Freiburg, Germany. bisse@med1.ukl.uni-freiburg.de
Clin Biochem ; 40(9-10): 739-43, 2007 Jun.
Article em En | MEDLINE | ID: mdl-17467680
ABSTRACT

OBJECTIVE:

To evaluate the clinical use of Homocysteine-Primavette, a new blood collection medium for total homocysteine (tHcy) assay.

METHODS:

The agreement between baseline tHcy and tHcy in stabilized samples (40 h) was assessed for FPIA, HPLC, GC-MS, LC-MS, and ICL.

RESULTS:

tHcy concentrations in whole blood were stable for 40 h in Hcy-Primavette tubes.

CONCLUSION:

Primavette tubes are a good alternative for the accurate tHcy measurement and no readjustment of reference intervals is needed.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Coleta de Amostras Sanguíneas / Homocisteína Limite: Humans Idioma: En Ano de publicação: 2007 Tipo de documento: Article