Frame-based immobilization and targeting for stereotactic body radiation therapy.
Murray, Bryan; Forster, Kenneth; Timmerman, Robert.
Afiliação
  • Murray B; Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 75390, USA. bryan.murray@utsouthwestern.edu
Med Dosim ; 32(2): 86-91, 2007.
Article em En | MEDLINE | ID: mdl-17472887
ABSTRACT
Frame-based stereotactic body radiation therapy (SBRT), such as that conducted with Elekta's Stereotactic Body Frame, can provide an extra measure of precision in the delivery of radiation to extracranial targets, and facilitates secure patient immobilization. In this paper, we review the steps involved in optimal use of an extra-cranial immobilization device for SBRT treatments. Our approach to using frame-based SBRT consists of 4

steps:

patient immobilization, tumor and organ motion control, treatment/planning correlation, and daily targeting with pretreatment quality assurance. Patient immobilization was achieved with the Vac-Loc bag, which uses styrofoam beads to conform to the patient's shape comfortably within the body frame. Organ and motion control was assessed under fluoroscopy and controlled via a frame-mounted abdominal pressure plate. The compression screw was tightened until the diaphragmatic excursion range was < 1 cm. Treatment planning was performed using the Philips Pinnacle 6.2b system. In this treatment process, a 20 to 30 noncoplanar beam arrangement was initially selected and an inverse beam weight optimization algorithm was applied. Those beams with low beam weights were removed, leaving a manageable number of beams for treatment delivery. After planning, daily targeting using computed tomography (CT) to verify x-, y-, and z-coordinates of the treatment isocenter were used as a measure of quality assurance. We found our daily setup variation typically averaged < 5 mm in all directions, which is comparable to other published studies on Stereotactic Body Frame. Treatment time ranged from 30 to 45 minutes. Results demonstrate that patients have experienced high rates of local control with acceptable rates of severe side effects - by virtue of the tightly constrained treatment fields. The body frame facilitated comfortable patient positioning and quality assurance checks of the tumor, in relation to another set of independent set of coordinates defined by the body frame fiducials. The ability to impose abdominal compression proved to be a simple way to reduce target and tissue motion. SBRT with Stereotactic Body Frame enables comfortable patient immobilization and facilitates repeated registering and re-registering of the patient to the frame. With the body frame, large-dose-per fraction treatment is possible for localized tumor deposits with the aim of attaining a more therapeutic result.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Restrição Física / Radiocirurgia Limite: Humans Idioma: En Ano de publicação: 2007 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Restrição Física / Radiocirurgia Limite: Humans Idioma: En Ano de publicação: 2007 Tipo de documento: Article