[Phantom studies using echo contrast media to improve the Doppler color sonographic imaging of the superficial femoral artery in the adductor canal]. / Phantomuntersuchungen mit echokontrastgebenden Substanzen zur Verbesserung der farb-doppler-sonographischen Darstellung der A. femoralis superficialis im Adduktorenkanal.
Rofo
; 154(1): 34-8, 1991 Jan.
Article
em De
| MEDLINE
| ID: mdl-1846690
The adductor canal was simulated using 2.6 cm muscular tissue and 2 fasciae to analyse the limits of colour-coded Doppler sonography (angiodynography) in this region. Defects in the spectral signal cause a significant underestimation of mean, peak systolic and peak diastolic (backflow) velocities and of calculated blood flow. Furthermore the pulsatility index is overestimated and the colour-coded visualisation of the arteries is almost lost. For the most part, these changes can be compensated by administration of a sonographic contrast agent (SH U 454). A minimum of 9 mg microbubbles/ml blood is required. Nevertheless, the adjustment of system controls (e.g. transducer power) becomes more difficult and an ideal setting impossible.