Electromagnetic interference in an implantable loop recorder caused by a portable digital media player.
Thaker, Jay P; Patel, Mehul B; Jongnarangsin, Krit; Liepa, Valdis V; Castellani, Mark; Thakur, Ranjan K.
Afiliação
  • Thaker JP; Sparrow Health System, Michigan State University, Lansing, Michigan, USA.
Pacing Clin Electrophysiol ; 31(10): 1345-7, 2008 Oct.
Article em En | MEDLINE | ID: mdl-18811818
The implantable loop recorder has been shown to be a cost-effective tool for diagnosis of intermittent cardiovascular symptoms such as syncope and palpitations. Electromagnetic interference in these recorders may be caused by commonly encountered electronic devices such as antitheft electronic surveillance systems and magnetic resonance imaging cameras. In this report, we describe interference in two patients with implantable loop recorders from a portable digital media player.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses e Implantes / Processamento de Sinais Assistido por Computador / Eletrocardiografia Ambulatorial / Falha de Equipamento Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses e Implantes / Processamento de Sinais Assistido por Computador / Eletrocardiografia Ambulatorial / Falha de Equipamento Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2008 Tipo de documento: Article