Your browser doesn't support javascript.
loading
Visual and refractive outcomes of LASIK with the SCHWIND ESIRIS and WaveLight ALLEGRETTO WAVE Eye-q excimer lasers: a prospective, contralateral study.
Mearza, Ali A; Muhtaseb, Mohammed; Aslanides, Ioannis M.
Afiliação
  • Mearza AA; Emmetropia Mediterranean Eye Institute, Heraklion, Crete, Greece. ali@eyedoc.eu.com
J Refract Surg ; 24(9): 885-90, 2008 11.
Article em En | MEDLINE | ID: mdl-19044228
PURPOSE: To compare the safety, efficacy, and predictability of LASIK with the SCHWIND ESIRIS and WaveLight ALLEGRETTO WAVE Eye-Q excimer laser platforms. METHODS: This prospective study comprised 44 eyes of 22 consecutive patients who were treated with LASIK using the Moria M2 microkeratome. One eye was treated with the SCHWIND ESIRIS laser and the fellow eye treated with the WaveLight ALLEGRETTO WAVE Eye-Q laser. All eyes operated with the SCHWIND ESIRIS were treated with standard aspheric ablation, whereas the eyes operated with the WaveLight ALLEGRETTO WAVE Eye-Q received treatment with three different ablation types according to the common practice at our clinic. Outcome measures were uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity (BSCVA), manifest refraction, and proximity to target refraction at 6-month follow-up. RESULTS: At 6 months postoperative, mean decimal UCVA was 0.96+/-0.22 (range: 0.3 to 1.2) for ESIRIS eyes and 0.98+/-0.17 (range: 0.6 to 1.2) for ALLEGRETTO eyes (P=.57). Mean postoperative spherical equivalent refraction was -0.02+/-0.28 diopters (D) (range: -0.75 to +0.75 D) for ESIRIS eyes and 0.11+/-0.91 D (range: -1.00 to +3.88 D) for ALLEGRETTO eyes (P=.49). Of the ESIRIS eyes, 20/22 (91%) were within +/-1.00 D of target refraction and 20/22 (91%) were within +/-0.50 D of target refraction. Of the ALLEGRETTO eyes, 20/22 (91%) and 19/22 (86%) were within +/-1.00 D and +/-0.50 D, respectively, of target refraction. No patient lost > or =2 lines of BSCVA in either group. CONCLUSIONS: No differences were seen in safety and efficacy outcome parameters between the SCHWIND ESIRIS and WaveLight ALLEGRETTO WAVE Eye-Q excimer lasers when used according to a previously established treatment algorithm at our clinic in the treatment of refractive error.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refração Ocular / Acuidade Visual / Ceratomileuse Assistida por Excimer Laser In Situ / Lasers de Excimer / Miopia Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Refração Ocular / Acuidade Visual / Ceratomileuse Assistida por Excimer Laser In Situ / Lasers de Excimer / Miopia Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2008 Tipo de documento: Article