A versatile high-resolution x-ray imager (HRXI) for laser-plasma experiments on OMEGA.
Rev Sci Instrum
; 79(10): 10E904, 2008 Oct.
Article
em En
| MEDLINE
| ID: mdl-19044559
A high-resolution x-ray imager (HRXI) devoted to laser-plasma experiments combines two state-of-the-art technologies developed in France: a high-resolution x-ray microscope and a high-speed x-ray streak camera. The resulting streaked imager achieves spatial and temporal resolutions of approximately 5 microm and approximately 10 ps, respectively. The HXRI has recorded enhanced spatial and temporal resolution radiographs of indirectly driven targets on OMEGA. This paper describes the main features of the instrument and details the activation process on OMEGA (particularly the alignment). Recent results obtained on joint CEA/LLE radiographic OMEGA experiments will also be presented.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2008
Tipo de documento:
Article