Limitations of 64-detector-row computed tomography coronary angiography: calcium and motion but not short experience.
Mir-Akbari, H; Ripsweden, J; Jensen, J; Pichler, P; Sylvén, C; Cederlund, K; Rück, A.
Afiliação
  • Mir-Akbari H; Division of Cardiology, Department of Internal Medicine, Karolinska Institute, Karolinska University Hospital, Stockholm, Sweden.
Acta Radiol ; 50(2): 174-80, 2009 Mar.
Article em En | MEDLINE | ID: mdl-19169918
ABSTRACT

BACKGROUND:

Recently, 64-detector-row computed tomography coronary angiography (CTA) has been introduced for the noninvasive diagnosis of coronary artery disease.

PURPOSE:

To evaluate the diagnostic capacity and limitations of a newly established CTA service. MATERIAL AND

METHODS:

In 101 outpatients with suspected coronary artery disease, 64-detector-row CTA (VCT Lightspeed 64; GE Healthcare, Milwaukee, Wisc., USA) was performed before invasive coronary angiography (ICA). The presence of >50% diameter coronary stenosis on CTA was rated by two radiologists recently trained in CTA, and separately by an experienced colleague. Diagnostic performance of CTA was calculated on segment, vessel, and patient levels, using ICA as a reference. Segments with a proximal reference diameter <2 mm or with stents were not analyzed.

RESULTS:

In 51 of 101 patients and 121 of 1280 segments, ICA detected coronary stenosis. In 274 of 1280 (21%) segments, CTA had non-diagnostic image quality, the main reasons being severe calcifications (49%), motion artifacts associated with high or irregular heart rate (45%), and low contrast opacification (14%). Significantly more women (43%) had non-diagnostic scans compared to men (20%). A heart rate above 60 beats per minute was associated with significantly more non-diagnostic patients (38% vs. 18%). In the 1006 diagnostic segments, CTA had a sensitivity of 78%, specificity of 95%, positive predictive value (PPV) of 54%, and negative predictive value (NPV) of 98% for detecting significant coronary stenosis. In 29 patients, CTA was non-diagnostic. In the remaining 72 patients, sensitivity was 100%, specificity 65%, PPV 79%, and NPV 100%. The use of a more experienced CTA reader did not improve diagnostic performance.

CONCLUSION:

CTA had a very high negative predictive value, but the number of non-diagnostic scans was high, especially in women. The main limitations were motion artifacts and vessel calcifications, while short experience in CTA did not influence the interpretation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Angiografia Coronária / Estenose Coronária Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2009 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Angiografia Coronária / Estenose Coronária Tipo de estudo: Diagnostic_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2009 Tipo de documento: Article