Comparison of VerifyNow-P2Y12 test and Flow Cytometry for monitoring individual platelet response to clopidogrel. What is the cut-off value for identifying patients who are low responders to clopidogrel therapy?
Godino, Cosmo; Mendolicchio, Loredana; Figini, Filippo; Latib, Azeem; Sharp, Andrew Sp; Cosgrave, John; Calori, Giliola; Cera, Michela; Chieffo, Alaide; Castelli, Alfredo; Maseri, Attilio; Ruggeri, Zaverio M; Colombo, Antonio.
Afiliação
  • Godino C; Department of Cardio-Thoracic and Vascular Diseases, San Raffaele Scientific Institute, Milan, Italy. cosmogodino@gmail.com.
Thromb J ; 7: 4, 2009 May 06.
Article em En | MEDLINE | ID: mdl-19419580
ABSTRACT

BACKGROUND:

Dual anti-platelet therapy with aspirin and a thienopyridine (DAT) is used to prevent stent thrombosis after percutaneous coronary intervention (PCI). Low response to clopidogrel therapy (LR) occurs, but laboratory tests have a controversial role in the identification of this condition.

METHODS:

We studied LR in patients with stable angina undergoing elective PCI, all on DAT for at least 7 days, by comparing 1) Flow cytometry (FC) to measure platelet membrane expression of P-selectin (CD62P) and PAC-1 binding following double stimulation with ADP and collagen type I either in the presence of prostaglandin (PG) E1; 2) VerifyNow-P2Y12 test, in which results are reported as absolute P2Y12-Reaction-Units (PRU) or % of inhibition (% inhibition).

RESULTS:

Thirty controls and 52 patients were analyzed. The median percentage of platelets exhibiting CD62P expression and PAC-1 binding by FC evaluation after stimulation in the presence of PG E1 was 25.4% (IQR 21.4-33.1%) and 3.5% (1.7-9.4%), respectively. Only 6 patients receiving DAT (11.5%) had both values above the 1st quartile of controls, and were defined as LR. Evaluation of the same patients with the VerifyNow-P2Y12 test revealed that the area under the receiver-operating-characteristic (ROC) curve was 0.94 (95% CI 0.84-0.98, p < 0.0001) for % inhibition and 0.85 (0.72-0.93, p < 0.005) for PRU. Cut-off values of ≤ 15% inhibition or > 213 PRU gave the maximum accuracy for the detection of patients defined as having LR by FC.

CONCLUSION:

In conclusion our findings show that a cut-off value of ≤ 15% inhibition or > 213 PRU in the VerifyNow-P2Y12 test may provide the best accuracy for the identification of patients with LR.