Your browser doesn't support javascript.
loading
Novel use of a vascular plug to anchor an azygous vein ICD lead.
Bar-Cohen, Yaniv; Takao, Cheryl M; Wells, Winfield J; Saxon, Leslie A; Cesario, David A; Silka, Michael J.
Afiliação
  • Bar-Cohen Y; Division of Cardiology, Childrens Hospital Los Angeles, University of Southern California, Los Angeles, CA 90027, USA. ybarcohen@chla.usc.edu
J Cardiovasc Electrophysiol ; 21(1): 99-102, 2010 Jan.
Article em En | MEDLINE | ID: mdl-19549034
We describe the case of a young patient with severe hypertrophic cardiomyopathy and marginal defibrillation thresholds (DFTs) at implant of a standard transvenous implantable cardioverter-defibrillator (ICD) system. The patient subsequently experienced multiple failed ICD shocks during a prolonged episode of spontaneous ventricular tachycardia/fibrillation. Placement of a second single-coil shocking lead in the azygous vein resulted in acceptable DFTs, but the new lead migrated superiorly within hours of the procedure. To stabilize the lead position, a vascular plug was placed in the distal azygous vein, and the shocking lead screw was actively fixated to the meshwork of the device. Subsequent testing confirmed both adequate defibrillation and stable lead position.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cardiomiopatia Hipertrófica / Prótese Vascular / Desfibriladores Implantáveis / Implantação de Prótese / Eletrodos Implantados / Cardiomiopatias Tipo de estudo: Diagnostic_studies Limite: Adolescent / Female / Humans Idioma: En Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cardiomiopatia Hipertrófica / Prótese Vascular / Desfibriladores Implantáveis / Implantação de Prótese / Eletrodos Implantados / Cardiomiopatias Tipo de estudo: Diagnostic_studies Limite: Adolescent / Female / Humans Idioma: En Ano de publicação: 2010 Tipo de documento: Article