Flow-through integrating cavity absorption meter: experimental results.
Musser, Joseph A; Fry, Edward S; Gray, Deric J.
Afiliação
  • Musser JA; Stephen F. Austin State University, Department of Physics and Astronomy, P.O. Box 13044 SFA Station, Nacogdoches, Texas 75962-3044, USA. musserja@sfasu.edu
Appl Opt ; 48(19): 3596-602, 2009 Jul 01.
Article em En | MEDLINE | ID: mdl-19571914
We report experimental results from a flow-through integrating cavity absorption meter. The operating range of the device is from 0.004 m(-1) to over 80 m(-1) of absorption. Absorption coefficients have been measured with 8% or less change in the presence of over 200 m(-1) of scattering in the medium. The instrument signal has been shown to be independent of flow rate up to 20 liters/min and thus independent of turbulence. This large operational range along with the ability to measure absorption independently of adverse scattering affects allows the instrument to be utilized in a wide range of environmental conditions.