Your browser doesn't support javascript.
loading
Seizure detection with a commercially available bedside EEG monitor and the subhairline montage.
Young, G Bryan; Sharpe, Michael D; Savard, Martin; Al Thenayan, Eyad; Norton, Loretta; Davies-Schinkel, Corrine.
Afiliação
  • Young GB; Department of Clinical Neurological Sciences, The University of Western Ontario, Room B10-106, University Hospital, 339 Windermere Rd, London, Ontario, N6A 5A5, Canada. bryan.young@lhsc.on.ca
Neurocrit Care ; 11(3): 411-6, 2009 Dec.
Article em En | MEDLINE | ID: mdl-19585276
ABSTRACT

INTRODUCTION:

Availability of standard, continuous electroencephalography (cEEG) monitoring in ICU is very limited, although commercially available 4-channel modules are present in many ICUs. We investigated the sensitivity of such modules compared with the more complete monitoring with a standard EEG system.

METHODS:

Seventy patients at high risk of seizures in the medical-surgical intensive care unit and Epilepsy Monitoring Unit were recorded simultaneously for at least 24 h with a 4-channel commercial ICU bedside monitoring system (Datex-Ohmeda) with a subhairline montage and a standard EEG machine (XLTEK) using the international 10-20 system of electrode placement. Recordings were interpreted independently from each other.

RESULTS:

The 4-channel recordings demonstrated a sensitivity of 68 and 98% specificity for seizure detection, and a sensitivity of 39% and a specificity of 92% for detection of spikes and PLEDs.

CONCLUSIONS:

The 4-channel EEG module has limited but practical usefulness for seizure detection when standard cEEG monitoring is not available.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Eletroencefalografia / Epilepsia / Monitorização Fisiológica Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2009 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Eletroencefalografia / Epilepsia / Monitorização Fisiológica Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2009 Tipo de documento: Article