Seizure detection with a commercially available bedside EEG monitor and the subhairline montage.
Neurocrit Care
; 11(3): 411-6, 2009 Dec.
Article
em En
| MEDLINE
| ID: mdl-19585276
ABSTRACT
INTRODUCTION:
Availability of standard, continuous electroencephalography (cEEG) monitoring in ICU is very limited, although commercially available 4-channel modules are present in many ICUs. We investigated the sensitivity of such modules compared with the more complete monitoring with a standard EEG system.METHODS:
Seventy patients at high risk of seizures in the medical-surgical intensive care unit and Epilepsy Monitoring Unit were recorded simultaneously for at least 24 h with a 4-channel commercial ICU bedside monitoring system (Datex-Ohmeda) with a subhairline montage and a standard EEG machine (XLTEK) using the international 10-20 system of electrode placement. Recordings were interpreted independently from each other.RESULTS:
The 4-channel recordings demonstrated a sensitivity of 68 and 98% specificity for seizure detection, and a sensitivity of 39% and a specificity of 92% for detection of spikes and PLEDs.CONCLUSIONS:
The 4-channel EEG module has limited but practical usefulness for seizure detection when standard cEEG monitoring is not available.