Your browser doesn't support javascript.
loading
Photon beam quality variations of a flattening filter free linear accelerator.
Georg, Dietmar; Kragl, Gabriele; Wetterstedt, Sacha af; McCavana, Patrick; McClean, Brendan; Knöös, Tommy.
Afiliação
  • Georg D; Department of Radiotherapy, Division Medical Radiation Physics, Medical University of Vienna, AKH Vienna, 1090 Vienna, Austria. dietmar.georg@akhwien.at
Med Phys ; 37(1): 49-53, 2010 Jan.
Article em En | MEDLINE | ID: mdl-20175465
ABSTRACT

PURPOSE:

Recently, there has been an increasing interest in operating conventional linear accelerators without a flattening filter. The aim of this study was to determine beam quality variations as a function of off-axis ray angle for unflattened beams. In addition, a comparison was made with the off-axis energy variation in flattened beams.

METHODS:

Two Elekta Precise linear accelerators were modified in order to enable radiation delivery with and without the flattening filter in the beam line. At the Medical University Vienna (Vienna, Austria), half value layer (HVL) measurements were performed for 6 and 10 MV with an in-house developed device that can be easily mounted on the gantry. At St. Luke's Hospital (Dublin, Ireland), measurements were performed at 6 MV in narrow beam geometry with the gantry tilted around 270 degrees with pinhole collimators, an attenuator, and the chamber positioned on the table. All attenuation measurements were performed with ionization chambers and a buildup cap (2 mm brass) or a PMMA mini phantom (diameter 3 cm, measurement depth 2.5 cm).

RESULTS:

For flattened 6 and 10 MV photon beams from the Elekta linac the relative HVL(theta) varies by about 11% for an off-axis ray angle theta = 10 degrees. These results agree within +/- 2% with a previously proposed generic off-axis energy correction. For unflattened beams, the variation was less than 5% in the whole range of off-axis ray angles up to 10 degrees. The difference in relative HVL data was less than 1% for unflattened beams at 6 and 10 MV.

CONCLUSIONS:

Off-axis energy variation is rather small in unflattened beams and less than half the one for flattened beams. Thus, ignoring the effect of off-axis energy variation for dose calculations in unflattened beams can be clinically justified.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radiometria / Radioterapia de Alta Energia / Terapia com Prótons Tipo de estudo: Diagnostic_studies / Evaluation_studies Idioma: En Ano de publicação: 2010 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radiometria / Radioterapia de Alta Energia / Terapia com Prótons Tipo de estudo: Diagnostic_studies / Evaluation_studies Idioma: En Ano de publicação: 2010 Tipo de documento: Article