The Clinical Laboratory Standards Institute ([CLSI] formerly NCCLS) reference broth microdilution testing method (protocol M27-A3) was compared with a commercially available methods (Sensititre YeastOne(®)) by testing two quality control strains and 102 isolates of Candida sp. and Cryptococcus sp. against fluconazole, itraconazole, ketoconazole, posaconazole, voriconazole, flucytosin, amphotericin B and caspofungin. Minimal inhibitory concentrations (MIC) endpoints were determined after 24h of incubation for Sensititre YeastOne(®) and after 24 and 48 h for CLSI microdilution method. Essential agreements between methods vary from 70.6 to 92.2%. Categorical agreements vary from 94.1% for 5FC to 72.6% for AMB. Sensititre YeastOne(®) reading appears to be useful for avoiding very major errors and this confirms the interest of this method for evaluating new antifungals activity in vitro.